2.4 LCD display
The high-contrast TFT colour display enables the video image to be observed under
a large angle of view. The following messages are shown on the display:
15. Display of a ' ' during the stimulus presentation
16. Display of a ' ' if the patient response button is pressed.
17.
The crosshairs help to centre the eye, scale = 1 mm interval
18. Warning or error message
19. Display of left (OS) or right eye (OD)
15
16
17
18
19
2.5 Control panel
The control panel is made of a comfortable and hard-wearing rubber material. All
buttons are backlit with white light to make navigation easy in a darkened room.
The light sources can be switched off if required, except for the display brightness
setting.
20.
Turn the refractive lens holder in and out
21. Start examination
22. Display brightness setting
23. Position chin rest left, right, up, down
20
21
22
23
2.6 Connections
24. Mains switch
25. Fuse holder for two fuses T3.15 AL / 250 V
26. Mains connection
27. Ethernet port
28. Plug-in mains power unit
WARNING!
All externally connected devices must comply with the standards rel-
evant to safety.
24
25
26
27
28
2.7 Housing
The optical components and electronics are protected from light and dust by fi ve
housing covers. They can be removed for servicing with just a few actions. Once
the four screws in the back panel have been removed, the panel, hood and both IR
covers can be lifted out. The optical unit and electronic components of the Octopus
900 are now accessible
WARNING!
Always disconnect the appliance from the mains power supply by pull-
ing out the mains cable before opening the appliance. Housing compo-
nents may be removed only by correspondingly trained and authorised
skilled personnel.
2.8 Cupola
The cupola of the Octopus 900 has a diameter of 600 mm and thus conforms to the
Goldmann standard.
2.9 Forehead rest
A wide, ergonomically designed forehead rest allows the patient to maintain a com-
fortable posture during the examination.