1. Safety
DANGER!
Failure to comply with these instructions may result in material damage
or pose a danger to patients or users.
WARNING!
These warnings must absolutely be complied with to guarantee
safe operation of the device and to avoid any danger to users and
to patients.
NOTE!
Important information: please read carefully.
1.1 Areas of application of the device
The device is intended to use in professional health care facility environment, like
doctor's practices, hospitals and optometrists and opticians premises, except near
of HF surgical equipment and RF shielded rooms of ME-systems for magnetic reso-
nance imaging. Some portable radio frequency equipment, like cell phones or RF
telephone equipment including antennas may interference medical devices. Such
equipment has to be kept in a distance of more than 30 cm (12 inches) from any
part of the instrument. Inobservance of this precaution may lower the correct func
-
tion of the instrument.
1.2 Patient population
The patient must be capable of sitting up straight and keeping his head still. He/she
must be physically and mentally able to cooperate well and is mentally capable of
following the examination. Patients must be at least 6 years old.
1.3 Ambient conditions
Transport:
Temperature
Air pressure
Relative humidity
from
from
from
−40°C
500 hPa
10%
to
to
to
+70°C
1060 hPa
95%
Storage:
Temperature
Air pressure
Relative humidity
from
from
from
−10°C
700 hPa
10%
to
to
to
+55°C
1060 hPa
95%
Use:
Temperature
Air pressure
Relative humidity
from
from
from
+10°C
800 hPa
30%
to
to
to
+35°C
1060 hPa
90%
Application height < 2,000 m above sea level
1.3 Shipment and unpacking
•
Before you unpack the appliance, check whether the packaging shows traces of
improper handling or damage. If it does, notify the transport company that deliv
-
ered the goods to you. Unpack the equipment together with a representative of
the transport company. Compile a report on any potentially damaged parts. This
report must be signed by you and by the representative of the transport company.
•
Leave the device in the packaging for a few hours before unpacking it (risk of con
-
densation).
•
Check the appliance for damage after it is unpacked. Return defective appliances
in the appropriate packaging.
•
Store the packaging material carefully so that it can be used for possible returns
or when moving.
1.4 Installation warnings
WARNING!
•
Do not modify this equipment without authorization of the manufacturer.
Installation and repairs may only be performed by trained specialists.
Any third-party device must be connected in compliance with the
EN 60601-1 standard.
Only original HS replacement parts may be used.
The device must not be stacked or placed in close proximity to other
electronic devices.
•
•
•
NOTE!
•The Octopus 900 must be placed in a completely darkened room.
•The use of accessories other than than those listed may result in
higher emissions or lower interference immunity of the Octopus 900
system.
•The software must be installed by trained personnel.
1.4 Operation and environment
DANGER!
Never use the device in potentially explosive environments where vola-
tile solvents (alcohol, petrol, etc.) and flammable anaesthetics are in
use.
WARNING!
The device must be switched off after every use. Otherwise there is a
risk of overheating when a protective dust cover is used.