3.3 Connect the network cable
Connect the Octopus 900 and PC with two Ethernet cables via the network switch
provided with the instrument. A computer network can also be connected via the
network switch. You will find further information in section 4 'Safe system configura
-
tion in accordance with EN 60601-1'.
3.4 Connect the electric power supply cable
The built-in mains power units operate with the voltages specified in the Technical
Data section. It is not necessary to select the voltage on the appliance. If a support
stand was also supplied, the Octopus 900 can be connected to the power socket in
the support stand's electrical connection box.
4. Safe system configuration in accordance
with EN 60601-1
4.1 System variant I, Octopus 900 with
laptop as control unit
NOTE!
A network switch (*) must always be interposed for perfect operation.
*
**
< 1.5 m **
Mains connection
LAN connection
Mains connection
safety isolating transformer
Ethernet cable LAN
(optional)
network switch
Laptop / Control unit
USB cable
Control unit
01-IFU_Octopus900-7220343_04070_eng.indd 10