Haag-Streit AG
Operation of C.MOR HD and C.MOR HD³ | 7
1500 4000089 02010_web
29
7.3.5 Choose White Balance Mode
In case of incorrect color reproduction, a white balance can be perfomed. The white balance func-
tion can be set to automatic or manual. With manual white balance, it is possible to adjust the val-
ues for the red and blue gain.
To change settings in the setup menu, the menu lock must be deactivated:
1. Open the
Setup Menu
.
2. Select the menu item
Miscellaneous
.
3. Set the value of the menu item
Menu Lock
to
OFF
.
ð
The red lock mark in the menu screen disappears.
4. Select the menu item
White Balance
.
5. Set
WB Mode
to the desired value:
–
AWB
: The white balance is adjusted automatically. To perform an automatic white balance,
Perform Automatic White Balance (AWB) [
. Unlike the auto-tracing white balance,
this setting is made only once when the function is triggered.
–
Manual
: The white balance can be set manually. This setting is carried out via
Manual
Gain Red
and
Manual Gain Blue
in the same menu.
By increasing the red and blue
gain values, these colors are displayed more vividly.
–
ATW
: Activates the auto-tracing white balance. In this mode, the color temperature is mon-
itored continuously and thereby white balance is set automatically. This function is recom-
mended when lighting conditions change frequently.
6. Confirm the entry by pressing the button [FUNCTION/SET] on the front panel of the CCU.
7. To activate the menu lock, set the value of the menu item
Menu Lock
to
ON
.
7.3.6 Adjust Automatic Sensitivity Settings
It is possible to carry out the automatic adjustment of the shutter and the light sensitivity (gain)
based on the image that is displayed on the monitor.
To use this function, the size of the measuring area must be set, which is to serve as the basis for
the automatic setting.
To change settings in the setup menu, the menu lock must be deactivated:
1. Open the
Setup Menu
.
2. Select the menu item
Miscellaneous
.
3. Set the value of the menu item
Menu Lock
to
OFF
.
ð
The red lock mark in the menu screen disappears.
4. Select the menu item
AE mode
.
5. Choose the
Area Select
value for the measuring area that is displayed on the monitor:
–
MID
: About 40% of the screen center is the measuring area.
–
WIDE
: About 60% of the screen center is the measuring area.
–
FULL
: The entire screen is the measuring area.
–
NARROW
: About 10% of the screen center is the measuring area.
–
USER
: Enables user defined setting.
6. Confirm the entry by pressing the button [FUNCTION/SET] on the front panel of the CCU.
7. To activate the menu lock, set the value of the menu item
Menu Lock
to
ON
.