BIOMED SERVICE MANUAL Electrosurgical Generator G400 Workstation
Model 777000
Gyrus ACMI G400 Workstation
BIOMED SERVICE MANUAL
Page 29 of 57
Part Number: 177079-AA
G400 Generator Test Results Page 2 of 3
Parameter
IEC
60601-1
IEC
60601-2-2
Limits(s)
Measured
Value
Protective Earth Impedance (PE)
18(f)
----
0.1 ohm (0.2
inc mains lead)
PE test current
18(f)
----
10 - 25A
A
Input voltage (V)
----
----
----
V
Input current: activating (I)
----
----
----
A
Input current: standby (I
Q
)
----
----
----
A
Input power (V x I)
7.1
----
440VA
VA
Earth leakage current (NC)
19.1(e)
----
500uA
uA
Earth leakage current (SFC)
19.2(a)
----
1mA
mA
Enclosure leakage current (NC)
19.1(d)
----
100uA
uA
Enclosure leakage current (SFC)
19.2(c)
----
500uA
uA
Patient leakage current (NC)
19.1(e)
----
100uA
uA
Patient auxiliary leakage current (NC)
19.1(f)
----
100uA
uA
Patient auxiliary leakage current (SFC)
19.1(f)
----
500uA
uA
Patient leakage (mains on applied part)
19.2(b)
----
5mA
mA
PK RF output power checks
----
----
----
----
Modality
Set Watt
Load
----
----
----
----
PK3
200W
68
(Bk – Rd)
----
50.2
170 – 220W
W
PK3
200W
150
(Bk – Rd)
----
50.2
162 – 223W
W
PK3
200W
1000
(Bk – Rd)
----
50.2
92 – 135W
W
PK3
100W
150
(Bk – Rd)
----
50.2
83 – 121W
W
DES
120W
68
(Bk – Rd)
----
50.2
92 – 135W
W
DES
120W
150
(Bk – Rd)
----
50.2
66 – 96W
W
DES
120W
1000
(Bk – Rd)
----
50.2
12.1 – 16.9W
W
DES
60W
68
(Bk – Rd)
----
50.2
51 – 76W
W
DES
60W
150
(Bk – Rd)
----
50.2
50 – 74W
W
DES
60W
1000
(Bk – Rd)
----
50.2
11.5 – 16.8W
W
PB1
100% CUT
200W
5
(Bk – Rd)
----
50.2
>80W
W
PB1
100% CUT
200W
500
(Bk – Rd)
----
50.2
157 – 211W
W
PB1
100% CUT
100W
5
(Bk – Rd)
----
50.2
57 – 76W
W
PB1
100% CUT
100W
500
(Bk – Rd)
----
50.2
85 – 115W
W
PBDES
200W
20
(Bk – Wh)
----
50.2
162 – 219W
W
PBDES
200W
300
(Bk – Wh)
----
50.2
32 – 47W
W
PBDES
100W
20
(Bk – Wh)
----
50.2
85 – 113W
W
PBDES
100W
300
(Bk – Wh)
----
50.2
32 – 47W
W