BIOMED SERVICE MANUAL Electrosurgical Generator G400 Workstation
Model 777000
Gyrus ACMI G400 Workstation
BIOMED SERVICE MANUAL
Page 14 of 57
Part Number: 177079-AA
9. Adjust the cut mode and wattage on the left side of the generator display to
PK3 200 (watts), set the ESU to 150 ohms input impedance and test for output.
While depressing the yellow footpedal, the output should measure between
179-216 watts.
10. If any questions or the generator does not meet the identified output
specification(s), call Gyrus ACMI Technical Services detailed on the contacts
page.
Test Mode
Load Impedance Ohms Limits (Watts)
DES 60W (COAG)
150
50-74
DES 120W (COAG)
150
66-96
PK3 100W (CUT)
150
85-115
PK3 200W (CUT)
150
179-216