Chapter 2: Electrical Safety Tests
LS 06311-002US Rev A
23
Table 4. Recommended Temperature Sensing Test:
TITLE
NOMINAL
TEST VALUE
PASS LIMIT
CHECK INSTRUCTION
Thermocouple
Check
Site Room
Ambient
The minimum
and maximum of
all 4 temperature
channels read by
the RF Generator
are within 4°C.
Click the message column to sort data and scroll
down to find “Hand Piece Connected TC’s” (Figure
2.7). If the maximum and minimum of the 4
temperature values are within a 4°C range, the
test is passing.
If the difference of maximum and minimum values is larger 4°C, contact Gynesonics for replacement RF
Generator unit.