Chapter 2: Electrical Safety Tests
LS 06311-002US Rev A
21
2.4
Patient Lead Leakage
Measure patient lead leakage using Patient Lead Grounding Cable ACCY-014. Conditions of test
include:
•
Power plug connected normally
•
System powered on
•
Ground switch CLOSED (NC limit
≤ 100 μA
), and separately with
ground switch OPEN (SFC limit
≤ 500 μA
)
•
Leakage current measured between all patient leads (RFA Handpiece and Dispersive
Electrodes) connected together and ground
Figure 2-5. Patient Lead Leakage Test Connection