6.1
Malfunctions
6.1.1
Equipment cannot be switched on
1. Check if the mains voltage has failed.
2. Check if the main switch is switched on (“I”).
3. Check if the power cord and the fuses are in order. If necessary,
replace the fuses. See
§6.1.2
.
4. Contact your dealer if the equipment still cannot be switched on.
6.1.2
Replacing the fuses
1. Switch the main switch off (“O”).
2. Unplug the power cord from the equipment.
3. Pull the fuse holder carefully out of the equipment. If necessary, use
a screwdriver.
4. Replace both fuses. If necessary, order new fuses from your dealer.
5. Install the fuse holder and plug in the power cord.
6. Switch the main switch on again (“I”).
6.1.3
Equipment does not react to commands
The safety system of the equipment has ascertained a fault. You cannot
continue to work.
1. Disconnect the connection to the patient.
2. Switch the main switch off (“O”).
3. Wait 5 seconds and switch the main switch on again (“I”).
4. Contact your dealer if the fault reappears.
Component
Problem
Solution
Vaco 200 unit
Equipment cannot be
switched on
See
§6.1.1
.
Equipment does not react to
commands
See
§6.1.3
.
Vacuum
electrodes
Contamination by ionization See
§6.1.4
.
Sponges
Furring
Replace the sponges
Bad conduction
Replace the sponges
OBJ_BUCH-62-001.book Page 20 Thursday, June 28, 2007 10:05 AM
Summary of Contents for VACO 200
Page 28: ...OBJ_BUCH 62 001 book Page 28 Thursday June 28 2007 10 05 AM...
Page 54: ...OBJ_BUCH 64 001 book Page 54 Wednesday June 27 2007 3 25 PM...
Page 80: ...OBJ_BUCH 66 001 book Page 80 Wednesday June 27 2007 3 38 PM...
Page 108: ...OBJ_BUCH 69 001 book Page 108 Wednesday June 27 2007 3 43 PM...
Page 134: ...OBJ_BUCH 65 001 book Page 134 Wednesday June 27 2007 3 48 PM...
Page 160: ...OBJ_BUCH 67 001 book Page 160 Wednesday June 27 2007 3 54 PM...
Page 186: ...Vaco 200 186 OBJ_BUCH 63 001 book Page 186 Thursday June 28 2007 10 32 AM...
Page 188: ...OBJ_BUCH 63 001 book Page 188 Thursday June 28 2007 10 32 AM...