The accuracy does not match the specification.
•
The display shows “fuse broken” message.
•
The display shows “output shorted” message.
•
The display shows “waiting for 10secs” message.
•
The display shows “OCP” message.
The accuracy does not match the specification.
Make sure the device is powered On for at least 30 minutes, within
+18°C~+28°C. This is necessary to stabilize the unit to match the
specification.
The display shows “fuse broken” message.
The internal protection fuse is open. Contact your local dealer or
GW Instek.
The display shows “output shorted” message.
Make sure the DUT is correctly connected to the APS-7000 output
terminals and that no shortcut exist. A DUT malfunction may also
be the cause.
The display shows “waiting for 10secs” message.
A protection function is likely in force, check the DUT settings.