PEL-3000 Series Parallel Assembly Guide
11
The current sink power of PEL-3111 and PEL-3211 is 1KW:2KW;
therefore, the ratio of current sink will always be the exactly 1:2.
E.g., PEL-3332 (3KW); when current sink is set 30A, Master (PEL-3111) will
sink current by 10A, whilst Booster (PEL-3211) will sink current by 20A.
E.g., PEL-3955 (9KW); when current sink is set 90A, Master (PEL-3111)
will sink current by 10A, whilst the rest of the 4 Booster (PEL-3211) will
sink current by 20A individually.
First dismantle all of the positive parallel copper sticks and parallel cables
in the rear output terminal of unit.
Check the 1st Booster current:
I.
Connect the parallel cable to the J2 connector in the rear of
PEL-3111 in one end and the J1 connector in the rear of PEL-3211
in the other end.
II.
Power on both PEL-3111 and the 1st PEL-3211, while keep the
rest units power-off. Set the unit PEL-3111 as the parallel mode
via the following procedure:
III.
Select from the main interface
a.
Configure > Next Menu > Parallel > Operation; select: Master
b.
Configure > Next Menu > Parallel > Parallel; select: OFF
c.
Configure > Next Menu > Parallel > Booster; select: 1