Gurin Accura Pulse AP-WBPM810 User Manual Download Page 4

CAUTION

* This device is intended for adult use in homes only. 

* The device is not suitable for use on neonatal patients, pregnant women,patients with 

implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, 

atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy 

or arterio-venous shunt or people who received a mastectomy. Please consult your doctor 

prior to using the unit if you suffer from illnesses.

* The device is not suitable for measuring the blood pressure of children. Ask your doctor 

before using it on older children.

* The device is not intended for patient transport outside a healthcare facility.

* The device is not intended for public use.

* This device is intended for no-invasive measuring and monitoring of arterial blood 

pressure. It is not intended for use on extremities other than the wrist or for functions other 

than obtaining a blood pressure measurement.

* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your 

blood pressure.Do not begin or end medical treatment without asking a physician for 

treatment advice.

* If you are taking medication,consult your physician to determine the most appropriate 

time to measure your blood pressure. Never change a prescribed medication without 

consulting your physician.

* Do not take any therapeutic measures on the basis of a self measurement. Never alter 

the dose of a medicine prescribed by a doctor. Consult your doctor if you have any 

question about your blood pressure.

* When the device was used to measure patients who have common arrhythmias such as 

atrial or ventricular premature beats or atrial fibrillation, the best result may occur with 

deviation. Please consult your physician about the result.

* Don't kink the connection tube during use, otherwise, the cuff pressure may continuously 

increase which can prevent blood flow and result in harmful injury to the PATIENT.

* When using this device, please pay attention to the following situation which may interrupt 

blood flow and influence blood circulation of the patient, thus cause harmful injury to the 

patient: connection tubing kinking too frequent and consecutive multiple measurements; 

the application of the cuff and its pressurization on any wrist where intravascular access or 

therapy, or an Arteriovenous (A-V) shunt, is present; inflating the cuff on the side of a 

mastectomy.

* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.

*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied 

around simultaneously, because this could cause temporary loss of function of those 

simultaneously-used monitoring ME equipment.

*On the rare occasion of a fault causing the cuff to remain fully inflated during 

measurement, open the cuff immediately. Prolonged high pressure (cuff pressure 

 

300mmHg or constant pressure 

 15mmHg for more than 3 minutes) applied to the wrist 

may lead to an ecchymosis.

*Please check that operation of the device does not result in prolonged impairment of 

patient blood circulation.

CAUTION

* When measurement, please avoid compression or restriction of the connection tubing.

* The device cannot be used with HF surgical equipment at the same time.

* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER 

was clinically investigated according to the requirements of ISO 81060-2:2013.

* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact 

the manufacturer.

* This device is contraindicated for any female who may be suspected of, or is pregnant. 

Besides providing inaccurate readings, the effects of this device on the fetus are unknown.

* Too frequent and consecutive measurements could cause disturbances in blood 

circulation and injuries.

* This unit is not suitable for continuous monitoring during medical emergencies or 

operations.Otherwise, the patient’s wrist and fingers will become anaesthetic, swollen and 

even purple due to a lack of blood.

* When not in use, store the device in a dry room and protect it against extreme moisture, 

heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.

* This device may be used only for the purpose described in this booklet. The manufacturer 

cannot be held liable for damage caused by incorrect application.

*This device comprises sensitive components and must be treated with caution. Observe 

the storage and operating conditions described in this booklet.

* The maximum temperature that the applied part can be achieved is 42.5

 while the 

environmental temperature is 40

.

* The equipment is not AP/APG equipment and not suitable for use in the presence of a 

flammable anesthetic mixture with air of with oxygen or nitrous oxide.

* Warning: No servicing/maintenance while the ME equipment is in use.

* The patient is an intended operator.

* The patient can measure data and change batteries under normal circumstances and 

maintain the device and its accessories according to the user manual.

* To avoid measurement errors, please avoid the condition of strong electromagnetic field 

radiated interference signal or electrical fast transient/burst signal.

* The blood pressure monitor and the cuff are suitable for use within the patient 

environment. If you are allergic to polyester, nylon or plastic, please don't use this device. 

* During use, the patient will be in contact with the cuff. The materials of the cuff have been 

tested and found to comply with requirements of ISO 10993-5:2009 and ISO 

10993-10:2010. It will not cause any potential sensization or irritation reaction.

* If you experience discomfort during a measurement, such as pain in the wrist or other 

complaints, press the START/STOP button to release the air immediately from the cuff. 

Loosen the cuff and remove it from your wrist.

* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. 

Should the cuff not deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff 

from the wrist and press the START/STOP button to stop inflation.

* Before use, make sure the device functions safely and is in proper working condition. 

Check the device, do not use the device if it is damaged in any way. The continuous use of 

a damaged unit may cause injury, improper results, or serious danger.

5

4

INTRODUCTION

INTRODUCTION

Summary of Contents for Accura Pulse AP-WBPM810

Page 1: ...M810 Please keep this manual well in order to reference in future Questions Warranty Support gurinproducts com Wrist Type Version 1 0 Distributed By Gurin Products LLC 2522 Chambers Road Suite 100 Tus...

Page 2: ...OOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure fluctuate throughout the day Why do I get a different bl...

Page 3: ...starts inflating the cuff Meanwhile the unit detects pressure oscillation generated by beat to beat pulsatile which is used to determine the systolic pressure and diastolic pressure as well as pulse r...

Page 4: ...d impairment of patient blood circulation CAUTION When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same tim...

Page 5: ...re between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equipment...

Page 6: ...ow happen shows Display is dim Display does not light up CUFF Type BF applied part BATTERY COMPARTMENT CAUTION Do not use new and used batteries together Do not use different types of batteries togeth...

Page 7: ...using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory year 2014 2054 time format 24 H 12 H 1 2 Press the button to change the TIME FORMAT b...

Page 8: ...loose the measurement will not be accurate Sit comfortably with your tested wrist resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your pa...

Page 9: ...1 2 The date and time will display alternately The corresponding time is 9 10 The corresponding date is May 11 When the monitor is off press button to show the average value of the latest three measu...

Page 10: ...signed to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so on and the last record 60 is dropped from the list CAUTION 3 If you want to check the other user s measu...

Page 11: ...oblems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of AccuraPulse Don t open or repair the device by yourself Please report to AccuraPulse if any unex...

Page 12: ...asurement position so please take the measurement under the same conditions 2 If the person takes medicine the pressure will vary more 3 Wait at least 3 minutes for another measurement Why do I get a...

Page 13: ...r assistance Refer to the warranty for contact information and return instructions Power supply Display mode Measurement mode Oscillographic testing mode Measurement range Accuracy Normal working cond...

Page 14: ...or basic safety and essential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment IEC EN 60601 1 2 2007...

Page 15: ...buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity IMMUNITY test 6 kV contact 8 kV air 6 kV contact 8 kV air 2 kV for power supply lines Not applicab...

Page 16: ...fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above th...