LCD Display Signal
* Do not wash the cuff in a washing machine or dishwasher!
* The service life of the cuff may vary by the frequency of washing, skin condition, and
storage state. The typical service life is 10000 times.
* It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of the
cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT
according to the local guidelines.
* Manufacturer will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions,etc., to assist to service personnel in parts repair.
* The operator shall not touch output of batteries and the patient simultaneously.
* Cleaning :Dust environment may affect the performance of the unit. Please use the soft
cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile
cleaners.
* The device doesn’t need to be calibrated within two years of reliable service.
* If you have any problems with this device, such as setting up, maintaining or using,
please contact the SERVICE PERSONNEL of AccuraPulse. Don’t open or repair the
device by yourself in the event of malfunctions. The device must only be serviced, repaired
and opened by individuals at authorized sales/service centers.
* Please report to AccuraPulse if any unexpected operation or events occur.
* Keep the unit out of reach of infants, young children or pets to avoid inhalation or
swallowing of small parts. It is dangerous or even fatal.
* Be careful to strangulation due to cables and hoses, particularly due to excessive length.
* At least 30 min required for ME equipment to warm from the minimum storage
temperature between uses until it is ready for intended use. At least 30 min required for ME
equipment to cool from the maximum storage temperature between uses until it is ready for
intended use.
* This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this
equipment and should be kept at least a distance d away from the equipment. The distance
d is calculated by the MANUFACTURER from the 80 MHz to 5.8 GHz column of Table 4
and Table 9 of IEC 60601-1-2:2007, as appropriate.
* Please use ACCESSORIES and detachable partes specified/ authorised by
MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the
user/patients.
* There is no luer lock connectors are used in the construction of tubing, there is a
possibility that they might be inadvertently connected to intravascular fluid systems,
allowing air to be pumped into a blood vessel.
* Please use the device under the environment which was provided in the user manual.
Otherwise, the performance and lifetime of the device will be impacted and reduced.
CAUTION
Systolic
blood pressure
High blood pressure
Diastolic
blood pressure
Low blood pressure
mmHg
Measurement unit the blood pressure
(1mmHg=0.133kPa)
Current time
Irregular heartbeat
Low battery
Batteries are low and need to be replaced.
SYMBOL
DESCRIPTION
EXPLANATION
Year/Month/Day,Hour/Minute
Heartbeat
User 1
User 2
Start measurement and save the
measuring results for user 1.
Start measurement and save the
measuring results for user 2.
Blood pressure monitor is detecting an
irregular heartbeat during measurement.
Blood pressure
level indicator
Indicate the blood pressure level
Blood pressure monitor is detecting a
heartbeat during measurement.
Motion indicator
Motion may result in an inaccurate
measurement.
7
6
INTRODUCTION
INTRODUCTION