LCD Display Signal

* Do not wash the cuff in a washing machine or dishwasher!

* The service life of the cuff may vary by the frequency of washing, skin condition, and 

storage state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of the 

cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT 

according to the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions,etc., to assist to service personnel in parts repair.

* The operator shall not touch output of batteries and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft 

cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile 

cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, 

please contact the SERVICE PERSONNEL of AccuraPulse. Don’t open or repair the 

device by yourself in the event of malfunctions. The device must only be serviced, repaired 

and opened by individuals at authorized sales/service centers.

* Please report to AccuraPulse if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or 

swallowing of small parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage 

temperature between uses until it is ready for intended use. At least 30 min required for ME 

equipment to cool from the maximum storage temperature between uses until it is ready for 

intended use.

* This equipment needs to be installed and put into service in accordance with the 

information provided in the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, walkie-talkies can affect this 

equipment and should be kept at least a distance d away from the equipment. The distance 

d is calculated by the MANUFACTURER from the 80 MHz to 5.8 GHz column of Table 4 

and Table 9 of IEC 60601-1-2:2007, as appropriate.

* Please use ACCESSORIES and detachable partes specified/ authorised by 

MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the 

user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a 

possibility that they might be inadvertently connected to intravascular fluid systems, 

allowing air to be pumped into a blood vessel.

* Please use the device under the environment which was provided in the user manual. 

Otherwise, the performance and lifetime of the device will be impacted and reduced.

CAUTION

Systolic 

blood pressure

High blood pressure 

Diastolic 

blood pressure

Low blood pressure

mmHg

Measurement unit the blood pressure 

(1mmHg=0.133kPa)

Current time

Irregular heartbeat

Low battery

Batteries are low and need to be replaced.

SYMBOL

DESCRIPTION

EXPLANATION

Year/Month/Day,Hour/Minute

Heartbeat

User 1

User 2

Start measurement and save the 

measuring results for user 1.
Start measurement and save the 

measuring results for user 2.
Blood pressure monitor is detecting an 

irregular heartbeat during measurement.

Blood pressure 

level indicator

Indicate the blood pressure level

Blood pressure monitor is detecting a 

heartbeat during measurement.

Motion indicator

Motion may result in an inaccurate 

measurement.

7

6

INTRODUCTION

INTRODUCTION

Summary of Contents for Accura Pulse AP-WBPM810

Page 1: ...M810 Please keep this manual well in order to reference in future Questions Warranty Support gurinproducts com Wrist Type Version 1 0 Distributed By Gurin Products LLC 2522 Chambers Road Suite 100 Tus...

Page 2: ...OOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure fluctuate throughout the day Why do I get a different bl...

Page 3: ...starts inflating the cuff Meanwhile the unit detects pressure oscillation generated by beat to beat pulsatile which is used to determine the systolic pressure and diastolic pressure as well as pulse r...

Page 4: ...d impairment of patient blood circulation CAUTION When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same tim...

Page 5: ...re between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equipment...

Page 6: ...ow happen shows Display is dim Display does not light up CUFF Type BF applied part BATTERY COMPARTMENT CAUTION Do not use new and used batteries together Do not use different types of batteries togeth...

Page 7: ...using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory year 2014 2054 time format 24 H 12 H 1 2 Press the button to change the TIME FORMAT b...

Page 8: ...loose the measurement will not be accurate Sit comfortably with your tested wrist resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your pa...

Page 9: ...1 2 The date and time will display alternately The corresponding time is 9 10 The corresponding date is May 11 When the monitor is off press button to show the average value of the latest three measu...

Page 10: ...signed to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so on and the last record 60 is dropped from the list CAUTION 3 If you want to check the other user s measu...

Page 11: ...oblems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of AccuraPulse Don t open or repair the device by yourself Please report to AccuraPulse if any unex...

Page 12: ...asurement position so please take the measurement under the same conditions 2 If the person takes medicine the pressure will vary more 3 Wait at least 3 minutes for another measurement Why do I get a...

Page 13: ...r assistance Refer to the warranty for contact information and return instructions Power supply Display mode Measurement mode Oscillographic testing mode Measurement range Accuracy Normal working cond...

Page 14: ...or basic safety and essential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment IEC EN 60601 1 2 2007...

Page 15: ...buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity IMMUNITY test 6 kV contact 8 kV air 6 kV contact 8 kV air 2 kV for power supply lines Not applicab...

Page 16: ...fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above th...