Battery powered mode: 

 2*AAA batteries (3V DC)
Blue LCD with backlight  V.A.35mm×46mm

Approx.106g (Excluding the batteries)
Approx.61mmx88mmx27.5mm (Exclude the cuff)
2*AAA batteries, user manual  

A10

This section includes a list of error messages and frequently 

asked questions for problems you may encounter with your wrist

blood pressure monitor. If the products not operating as you 

think it should, check here before arranging for servicing.

PROBLEM SYMPTOM CHECK THIS

REMEDY

No power

Low

batteries

Error

message

Display is dim or
will not light up.

Batteries are exhausted. Replace with new batteries.

Insert the batteries 
correctly.

Replace with new batteries.

Batteries are inserted

incorrectly.

Show on

the display Batteries are low.

Err 1 shows

The cuff is too loose.

Refasten the cuff and then

measure again.

Err 2 shows

The cuff is very tight

.

Refasten the cuff and then

measure again.

Err 3 shows

The pressure of the

cuff is excess.

Relax for a moment

and then measure again.

EExx, shows on 

the display.

A calibration error

occurred.

Retake the measurement.

If the problem persists,

contact the retailer or our

customer service 

department for further 

assistance. Refer to the 

warranty for contact 

information and return 

instructions.

Power supply
Display mode

Measurement mode

Oscillographic testing mode

Measurement range

Accuracy

Normal working condition

Storage & transportation

condition

Measurement perimeter

of the wrist

About 13.5cm-21.5cm

Weight

External dimensions

Attachment

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

Protection against 

ingress of water

Software version

Device classification

Internally Powered ME Equipment

WARNING: No modification of this equipment is allowed.

Pressure:

5°C-40°C within±3mmHg(0.4kPa) 

Pulse value: ±5%

Rated cuff pressure: 

0mmHg~299mmHg (0kPa ~ 39.9kPa)

Measurement pressure: 

SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa) 

DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)

Pulse value: (40-199)beat/minute

IP22: The first number 2: Protected against solid 

foreign objects of 12,5mm Ф and greater. The 

second number: Protected against vertically falling 

water drops when enclosure titled up to 15º. 

Vertically falling drops shall have no harmful 

effects when the enclosure is titled at any angle 

up to 15º on either side of the vertical.

Temperature: -20°C to 60°C

Relative humidity: 10%RH to 93%RH

Atmospheric pressure: 50kPa to 106kPa

Temperature: 5°C to 40°C

Relative humidity: ≤85%RH

Atmospheric pressure: 86kPa to 106kPa

23

22

SPECIFICATIONS

TROUBLESHOOTING

Summary of Contents for Accura Pulse AP-WBPM810

Page 1: ...M810 Please keep this manual well in order to reference in future Questions Warranty Support gurinproducts com Wrist Type Version 1 0 Distributed By Gurin Products LLC 2522 Chambers Road Suite 100 Tus...

Page 2: ...OOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure fluctuate throughout the day Why do I get a different bl...

Page 3: ...starts inflating the cuff Meanwhile the unit detects pressure oscillation generated by beat to beat pulsatile which is used to determine the systolic pressure and diastolic pressure as well as pulse r...

Page 4: ...d impairment of patient blood circulation CAUTION When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same tim...

Page 5: ...re between uses until it is ready for intended use At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equipment...

Page 6: ...ow happen shows Display is dim Display does not light up CUFF Type BF applied part BATTERY COMPARTMENT CAUTION Do not use new and used batteries together Do not use different types of batteries togeth...

Page 7: ...using your blood pressure monitor so that a time stamp can be assigned to each record that is stored in the memory year 2014 2054 time format 24 H 12 H 1 2 Press the button to change the TIME FORMAT b...

Page 8: ...loose the measurement will not be accurate Sit comfortably with your tested wrist resting on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your pa...

Page 9: ...1 2 The date and time will display alternately The corresponding time is 9 10 The corresponding date is May 11 When the monitor is off press button to show the average value of the latest three measu...

Page 10: ...signed to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so on and the last record 60 is dropped from the list CAUTION 3 If you want to check the other user s measu...

Page 11: ...oblems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of AccuraPulse Don t open or repair the device by yourself Please report to AccuraPulse if any unex...

Page 12: ...asurement position so please take the measurement under the same conditions 2 If the person takes medicine the pressure will vary more 3 Wait at least 3 minutes for another measurement Why do I get a...

Page 13: ...r assistance Refer to the warranty for contact information and return instructions Power supply Display mode Measurement mode Oscillographic testing mode Measurement range Accuracy Normal working cond...

Page 14: ...or basic safety and essential performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment IEC EN 60601 1 2 2007...

Page 15: ...buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity IMMUNITY test 6 kV contact 8 kV air 6 kV contact 8 kV air 2 kV for power supply lines Not applicab...

Page 16: ...fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above th...

Reviews: