Operator's guide
Depending on the digitizer type, sine-wave, step and broadband noise
signal generators may be available. In this section, broadband noise
calibration will be used to determine the complete sensor response in
one action. Please refer to the digitizer's manual for information on
other calibration methods.
1. In Scream!'s main window, right-click on the digitizer's icon and
select Control.... Open the
Calibration pane.
2. Select the calibration channel corresponding to the instrument,
and choose Broadband Noise. Select the component you wish to
calibrate, together with a suitable duration and amplitude, and
click Inject now. A new data stream, ending
C
n
(
n
= 0 – 7) or
MB
, should appear in Scream!'s main window containing the
returned calibration signal.
3. Open a Waveview window on the calibration signal and the
returned streams by selecting them and double-clicking. The
streams should display the calibration signal combined with the
November 2006
59