16
2.04
Safety functions
Important!
The emergency stop must only be used in case of failure/emergencies. The
supplier should be contacted before the lifting platform is put into service
again.
Description of the emergency stop function
If the lifting platform does not react to the control functions when it is in
motion – Press the emergency stop device. When the emergency stop has
been activated, all other functions of the lifting platform are put out of action.
Description of the safety list
function
A safety edge (pos. 16) has been
placed between the platform and
the bottom frame to eliminate the
risk of squeezing. If the safety
edge is activated during lowe-
ring, the lowering of the platform
will automatically stop and it will
be elevated a little. The lifting
platform cannot be lowered until
the control button “UP” has been
activated. If the safety edge is
activated when the platform is
not in motion, the platform will
remain on the level.on which it
was stopped. When activating the
control button “UP” the platform
will function again.
2.05
Connecting the charger/charging
Connection of charger
The charger must only be used in dry environments.
Connect the male plug of the charger to the power outlet.
Connect the other plug of the charger to the
battery box. The plug has been marked with
a label – 24 V DC.
Charging
The control panel of the steering handle
has been fitted with a green, a yellow and
a red indicator light. If the green indicator light
is switched on, the batteries are fully charged.
If the yellow or red indicator lights are
switched on 10 to 15 seconds after a lift,
the lifting platform should be recharged.