18
4.04
What maintenance is required by the owner?
Lubricate the lifting
platform with TRI-FLOW
or a similar lubricant
every six month or when
needed. The parts to
be lubricated are the
brakes, wires and
curve plates.
(Illustration of pos.22).
5.00
Service and Useful Life
5.01
Useful Life and Servicing
The LP12 must be inspected in accordance with national law, and at least
once a year. The service shall be carried out by a qualified service engineer
or the Guldmann Service Team.
The LP12 has an expected life span of 15 years. The LP12 must then be
assessed by qualified service staff as to its future use.
Spare parts lists and drawings are available from manufacturer or supplier.
5.02
Battery and Fuses
Change of batteries and fuses should be carried out by a service engineer or
by the Guldmann Service Team.
5.03
Inspection – All-round
During the inspection a service report shall be made, describing what has
been checked and exchanged. Worn or defective parts shall be replaced by
new spareparts from V. Guldmann A/S.
1. Visual check of the product
• Check the product for wear and tear
• Check the product for any deformity
• Check that the product does not show any other deficiencies
2. Test of the product, as during normal use
• Check all functions on the product, with and without load
(eg. Up, down, forward, backward, out and in)
• Check the emergency lowering is operational
• Check the emergency stop is operational
• Check the charging indicator is operational