6
1. Installation of the device must comply with appropriate standards
and associated electric safety requirements.
2. Never install the device in an explosive atmosphere and the
device must not be operated in an area with flammable gases
(anaesthetic mixtures, oxygen, etc.).
3. The installation site should prevent the device from being
impacted, and avoid the device from being splashed by water or
other liquids.
4. Do not install the device near or above the heat source. It must be
installed in a place where the surrounding air is sufficiently
circulated. There should be enough space around the device.
5. Do not place the device in direct sunlight or ultraviolet light.
6. This device can be moved, but please handle it with care.
7. Make sure the connection parts are dry before connecting the
wires to the unit. If necessary, blow it to dry with an air gun.
3.2 See the packing list for the machine configuration
The device consists of main unit, motor, motor tail, contra-angle
handpiece, power adapter.
Figure 1
3.3 Front View of the Main Unit