5
Chuck
Mechanical type chuck
Gear ratio
1:5 speed-rising
Shanks for rotary
instruments
Type3,φ1.6mm bur ( ISO 1797-1-2011)
Overall length of rotary
instrument
19mm
Maximum overall
length of
rotary instrument
21mm
Maximum working
diameter
of the rotary
instruments
2mm
Rod
A cylinder with rounded or chamfered ends
Minimum length of rod
12.5mm
Cooling water pressure
0.5-2 bar(1.5 bar is recommended)
Cooling air pressure
1.5-6 bar
Cooling air
consumption
under pressure of 2 bar
> 1.5 Nl/min
Bur extraction force
≥24N
2.3 Device safety classification
2.3.1 Type of protection against electric shock: Class II equipment
2.3.2 Degree of protection against electric shock: Type B applied
part
2.3.3 Degree of main unit protection against harmful ingress of
water: Ordinary equipment (IPX0). Not waterproof.
2.3.4 Classified by operation mode: Continuous operating device
2.3.5 Degree of safety application in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide: Equipment
cannot be used in the presence of a flammable anesthetic mixture
with air, oxygen, or nitrous oxide.
3 Product structure and composition
3.1 Safety requirements during installation