Steps
Motor
Contra-angle handpiece
Maintenance
1. Visual inspection: Before packaging and
sterilization, make sure that the motor has
been maintained according to manufacturer's
instruction. Visual inspection for cleanliness of
the motor.
Check the motor:
1. If there is still visible stain on the motor after
cleaning, the entire cleaning process must be
repeated.
If the motor is obviously damaged, smashed,
detached, corroded or bent, it must be discarded
and not allowed to continue to be used.
1. Functional Test and Visual
inspection
Visually inspect the cleanliness
of the Handpiece, and then
reassemble it. Perform Functional
test according to instructions of
use. If there is still visible stain on
the device after cleaning, the entire
cleaning process must be repeated.
Before packaging and sterilization,
make sure that the contra-angle
handpiece has been maintained
acc. to manufacturer's instruction.
If the device is obviously damaged,
smashed, detached, corroded or
bent, it must be discarded and not
allowed to continue to be used.
If the accessories are found to be
damaged, please replace it before
use. And the new accessories for
replacement must be cleaned,
disinfected and dried.
Use of an 510(k) cleared lubricant
(i.e. NSK PANA-SPRAY) to
lubricate the handpiece and dried
it prior to sterilization. Aim the
nozzle of lubricant bottle to the
air hole at the end of the contra-
angle handpiece to inject oil for 1-2
seconds.