2.7 In order to confirm the file clip and measuring wire makes good contact, test the
wire connecting before each use(See 3.1.3).
2.8 The file clip, lip hook and touch probe are reusable. Please make sure they are
autoclaved under high pressure and high temperature before each operation. The
endo files should not be used more than 3 times.
2.10 When the battery is low, a battery indicator "
" appears on the status bar
of the screen. Please recharge the battery when the battery charging prompt icon is
blinking.
2.11 Please use original components, the components made by other companies may
cause inaccurate measurement or un-measurable.
2.12 Avoid the connection between the outside and inside liquid of endodontic during
measuring in order to avoid the measuring difference.
2.13 Keep endo file and file clip away from any other metal or instruments.
2.14 To ensure that short circuits do not impair the measurements, be particularly
careful with patients fitted with metal crowns or bridges. Please confirm the wetness
of the endo to ensure the reliability of the measuring. If it is confirmed that the endo
file hasn't reached the apex yet the data showed on the apex locator is too low, please
check whether the endo is too dry and confirm it with X-ray.
2.15 This device have electromagnetic interference, the patient or doctor who with
a heart pace maker are forbidden to use this device and the device is susceptible