9
b) Make sure if the plug of the measuring wire is inserted into the socket correctly.
c) Make sure if the file clip and lip hook are connected well to the measuring wire.
d) Make the file clip nip the lip hook[as showed in picture 4], the “Connection
test
” icon must appear on the status bar[as showed in picture 5]. If no icon
appears, it means that the file clip or the measuring wire was damaged, should be
replaced.
3.1.4 Explanations on the interfaces displayed
a) The screen displays the front region of the apical foramen by bars. Please refer to
the blue region as showed. [Picture 6(a)]
b) The file has gone to the position near by the apical foramen when the yellow bars
displayed [Picture 6 (b)].
c) The file has exceeded the apical foramen when the "OVER" bars displayed. A
continuous beep sound will be generated at the same time [Picture 6 (c)].
3.1.5 Enlarged display of apical foramen. [Picture 7]
Picture 6
(a)
Picture 6
(b)
Picture 6
(c)