1.6.3 Measurement of the tooth length of transplantation and retransplantation.
1.7 Contraindication
We do not advise the use of Minipex on patients fitted with pacemakers (or other
electrical equipment) or on those patients who are advised not to use the electric
equipment (like electric shaver, electric blower) for safety reasons.
1.8 The classification of the device
1.8.1 Type of protection against electric shock: Class
Ⅱ
equipment
1.8.2 Degree of protection against electric shock: Type BF applied part
1.8.3 Degree of protection against water shock: Ordinary equipment ( IPX0 )
1.8.4 Device not suitable for being used in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
1.8.5 Operation mode: Continous operation
1.9 The main technical specifications
1.9.1 Battery: 3.7V/320mAh
1.9.2 Adapter: ~100V-240V 50Hz/60Hz
1.9.3 Consumption power: ≤0.5W
1.9.4 Screen: Segment digital tube
1.9.5 Buzzer alert: The buzzer will alert when the endo file is less than 2mm to the
apex.
1.9.6 Operation condition
a) Environment temperature: +5
℃~
+40
℃