vacuum process) can be used for sterilization, and other sterilization
procedures are prohibited:
4.6.1 The steam sterilizer complies with EN13060 or is certified by EN
285 and complies with EN ISO 17665;
4.6.2 The maximum sterilization temperature is 138℃ ;
4.6.3 At a temperature of 134° C, the sterilization time is at least 5
minutes;
4.6.4 The maximum sterilization time at 134° C is 20 minutes.
5. Storage and transportation
5.1 Storage
5.1.1 The product should be handled with care and away from the
earthquake source. Ensure that it is installed or stored in a cool, dry and
ventilated place indoors.
5.1.2 Do not mix the product with toxic, corrosive, flammable, or
explosive materials during storage.
5.1.3 The product should be stored in the environment with relative
humidity of 10% ~ 90%, atmospheric pressure of 50kPa ~ 106kPa and
temperature of -10℃~ +40℃ .
5.1.4 After sterilization, the product should be packaged in a medical