be sterilized, the powder storage chamber, tail cord connector and the
connecting end between the two cannot be sterilized;
② The product is designed for a large number of sterilization cycles,
and the materials for manufacturing are selected accordingly. However,
thermal and chemical stresses can lead to product aging each time they
are re-prepared for use. The maximum allowable disinfection times of
handpiece and spray-head are 1000.
4.1 Pre-treatment
4.1.1 Processing principles
Only after effective cleaning and disinfection can effective sterilization
be carried out. Please ensure that as part of the sterility of the product
in use, only fully validated equipment and product-specific procedures
are used for cleaning / disinfection and sterilization, and that validated
parameters are followed during each cycle.
Please also comply with the applicable legal requirements of your
country/region and the health regulations of hospitals or clinics, especially
other requirements regarding virus inactivation.
4.1.2 Preparation before cleaning
Tools: Air gun, tray, soft brush, soft cloth for cleaning and drying.
Processing must be carried out immediately after the operation, no later