maintenance” section) and packaging (see the “Packaging” section).
Dry the handpiece repeatedly if necessary (see the “Drying” section).
[Note]:
a) Disinfection must be carried out within 2 hours after the cleaning
phase. Automatic disinfection is preferred if conditions permit.
b) The washer-disinfector shall be CE certified to be effective according
to EN ISO 15883.
Conduct disinfection under high temperature. The temperature shall not
exceed 134 ° C and disinfection at this temperature shall not exceed 20
minutes.
c )The disinfection cycle shall comply with EN ISO 15883.
4.4 Drying
If the automatic drying function is not included in your cleaning and
disinfection process, please perform drying after cleaning and disinfection.
4.4.1 Lay a clean white cloth on the workbench, align the handpiece and
spray-head with the cloth, and then blow the handpiece and spray-head
with filtered dry compressed air (maximum pressure of 3 bar) for drying. If
there is no liquid sprayed on the white cloth, the drying is completed.
4.4.2 Drying can also be done directly in a medical drying cabinet (or
oven). The recommended drying temperature is 80 ℃~ 120 ℃ and the