12
EN ISO 15883. For the specific procedure, please refer to the automated
disinfection section in the next section “Disinfection”.
Notes:
a) The cleaning agent does not have to be pure water. It can be distilled
water, deionized water or multi-enzyme. But please ensure that the
selected cleaning agent is compatible with the product.
b) In washing stage, the water temperature should not exceed 45 °C,
otherwise the protein will solidify and it would be difficult to remove.
c) After cleaning, the chemical residue should be less than 10mg / L.
8.4 Disinfection
Disinfection must be performed no later than 2 hours after the cleaning
phase. Automated disinfection is preferred if conditions permit.
8.4.1 Automated disinfection-Washer-disinfector
·The Washer-disinfector complies with EN ISO 15883.
·Use high temperature disinfection function. The temperature does not
exceed 134 ° C, and the disinfection under the temperature cannot exceed
20 minutes.
·The disinfection cycle is in accordance with the disinfection cycle in
EN ISO 15883.
Cleaning and disinfecting steps by using Washer-disinfector
1. Carefully place the product into the disinfection basket. Fixation of
product is needed only when the product is removable in the device. The
products are not allowed to contact each other.
2. Start the program.
3. After the program is finished, remove the product from the washer-
disinfector, inspect (refer to section “Inspection and Maintenance”) and
packaging (refer to chapter “Packaging”). Dry the product repeatedly if
necessary (refer to section“Drying”).
The intrinsic suitability of the product for effective cleaning and
disinfection using the above automated cleaning and disinfection
procedures was verified by a certified facility.
Notes:
a) Before use, you must carefully read the operating instructions
provided by the equipment manufacturer to familiarize yourself with the
disinfection process and precautions.
b) With this equipment, cleaning, disinfection and drying will be
carried out together.
c) Cleaning: (c1) The cleaning procedure should be suitable for the
Summary of Contents for 28-780
Page 1: ......
Page 25: ...ZMN WI 09 459 V1 1 20200102 ...