![Guildline 7520 Operation And Instruction Manual Download Page 49](http://html.mh-extra.com/html/guildline/7520/7520_operation-and-instruction-manual_2278480049.webp)
Section 6
OM7520-B-00
27 September, 2019
6-2
6.2.
REMOTE COMMANDS
A brief description of each of the possible remote commands and their syntax in BNF
(Backus Naur Form) follows:
- words inside angle brackets (i.e. < and > ) are defined items
- :== means "is defined to be"
- | means "or"
- [] means optional
- required letters are shown in upper case but may be upper or lower case.
<digit> :== 0|1|2|3|4|5|6|7|8|9600
<letter>
:== A|B|C|...|Z|a|b|c|...|z
<string>
:== <letter> | <letter><string>
<boolean>
:== 0|1
<unsigned>
:== <digit> | <digit><unsigned>
<nr1>
:== [+|-]<unsigned>
<nr3>
:== <nr1>[.[<unsigned>]][E<nr1>]
<?>
:== <letter> | <digit>
<*>
:== <?> | [<*>]
: not to be confused with *
<DD>
:== <unsigned>
: limited to range 1...31
<MM>
:== <unsigned>
: limited to range 1...12
<YYYY>
:== <unsigned>
: limited to ranges 1999 and up
COMMAND
COMMENT
*IDN?
query, display identity of unit
*RST
reset the instrument to a known defined state
CONFigure:DIVider <ratio>
set the voltage ratio configuration
<ratio>:== 0|1|2|3|4
0=open
1=1:1
2=10:1
3=100:1
4=1000:1
Note: always go thru “OPEN” when changing ratio
CONFigure:DIVider?
query, display the ratio range