Section 4
OM7520-B-00
27 September, 2019
4-3
4.5.
COMPARISON MEASUREMENT CONFIGURATIONS
The Model 7520 Precision Voltage Divider can
be configured to allow relative and direct-
ratio measurements of signals input to the front panel terminations for each of the voltage
configurations shown in the diagrams below. These configurations are for forward and
reverse polarity measurements. In all configurations the 10 V reference is connected to the
Voltage Reference terminals and the Unit Under Test (i.e. UUT) is connected to the Input
Voltage. The VD box refers to the internal divider section.
A
B
C
D
E
F
Figure 4-1: 7520 Comparison Measurement Configurations
Note: In all configurations the output of the divider chain is connected to the high side
(red terminal) of the null detector.
10V
Reference
Standard
VD
IN
OUT
1:1000
Null Detector
UUT
0.01V
10V
0.01V
0.01V
10V
Reference
Standard
VD
IN
OUT
1:100
Null Detector
UUT
0.1V
10V
0.1V
0.1V
10V
Reference
Standard
VD
IN
OUT
1:10
Null Detector
UUT
1V
10V
1V
1V
10V
Reference
Standard
Null Detector
VD
Out
In
10:1
UUT
100V
10V
10V
100V
10V
Reference
Standard
Null Detector
VD
IN
OUT
1:1
UUT
10V
10V
10V
10V
10V
Reference
Standard
Null Detector
VD
Out
In
100:1
UUT
1000V
10V
10V
1000V