Section 3
OM7520-B-00
27 September, 2019
3-21
NOTE: In this configuration the Output Terminal is not used and should be
disconnected from any other equipment. Leaving this terminal connected to any
other device may result in loading errors.
NOTE: all 6 configurations are covered with the above shown connections.
(Reference Section 4 Configurations A, B, C, D, E, and F).
3.11.
0.01 VOLT, 1:1000 RATIO COMPARISON
Configuration A is used for the comparison of a 0.01 volt source to a 10 volt reference
standard. Connect the measurement components as described in Figure 3-5.
Configure the touch screen in the “Comparator” menu (section 3.4.3) for the 10 mV
comparison and enable the divider. Turn on the output from the 10 volt reference
connected to the Voltage Reference terminals and the output from the 0.01 volt source
connected to the Voltage Input terminals and adjust the 0.01 volt source for a null
indication on the null detector. Allow sufficient time for the measurement to stabilize.
Once a null is obtained the 0.01 volt source is equal to the 10 volt reference standard
certified value divided by the 1:1000 ratio of the 7520.
3.12.
0.1 VOLT, 1:100 RATIO COMPARISON
Configuration B is used for the comparison of a 0.1 volt source to a 10 volt reference
standard. Connect the measurement components as described in Figure 3-5.
Configure the touch screen in the “Comparator” menu (section 3.4.3) for the 100mV
comparison and enable the divider. Turn on the output from the 10 volt reference
connected to the Voltage Reference terminals and the output from the 0.1 volt source
connected to the Voltage Input terminals and adjust the 0.1 volt source for a null
indication on the null detector. Allow sufficient time for the measurement to stabilize.
Once a null is obtained the 0.1 volt source is equal to the 10 volt reference standard
certified value divided by the 1:100 ratio of the 7520.
3.13.
1 VOLT, 1:10 RATIO COMPARISON
Configuration C is used for the comparison of a 1 volt source to a 10 volt reference
standard. Connect the measurement components as described in Figure 3-5.
Configure the touch screen in the “Comparator” menu (section 3.4.3) for the 1V
comparison and enable the divider. Turn on the output from the 10 volt reference
connected to the Voltage Reference terminals and the output from the 1 volt source
connected to the Voltage Input terminals and adjust the 1 volt source for a null indication
on the null detector. Allow sufficient time for the measurement to stabilize. Once a null
is obtained the 1 volt source is equal to the 10 volt reference standard certified value
divided by the 1:10 ratio of the 7520.