background image

H49.0.02.6C-04

Operating Manual GMH 3750

Page 3

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   

                                                                                                                                                                                                                                       

1.2  Operating And Maintenance

 

Battery Operation

The battery has been used up and needs to be replaced, if

and „bAt“ are shown in lower display.

The device will, however, continue operating correctly for a certain time.
The battery has been completely used up, if ´bAt´ is shown in the upper display.
The battery has to be taken out, when storing device above 50°C.

     Hint: We recommend to remove the battery if device is not used for a longer period of time!

 

Mains Operation

Attention:

 When using a power supply unit please note that operating voltage has to be 10.5 to 12 V DC. Do not

apply overvoltage!! Simple 12V-power supplies often have excessive no-load voltage. We, therefore, recommend
using regulated voltage power supplies. Trouble-free operation is guaranteed by our power supply GNG10/3000.
Prior to connecting the plug power supply with the mains supply make sure that the operating voltage stated at the
power supply is identical to the mains voltage.

  Treat device and probes carefully. Use only in accordance with above specification. (do not throw, hit against etc.).

Protect plugs and sockets from soiling.

  To disconnect sensor plug do not pull at the cable but at the plug.

  When connecting the probe the plug will slide in smoothly if plug is entered correctly.

 

Selection of Output-Mode

: The output can be used as serial interface or as analogue output. This choice has to

be done in the configuration menu.

1.3 Connections

1.

Output:

 Operation as interface: Connect to optically isolated

interface adapter (accessory: GRS 3100 or GRS3105)
Operation as analogue output: Connection via suitable cable.
Attention: The output mode has to be configured (p.r.t. 2.7) and
influences battery life!

2.

Probe connection 

Pt100 4-wire

3.

The 

mains socket

 is located at the left side of the instrument

1.4 Display Elements

1 =  Main Display:

Currently measured temperature

2 =  Auxiliary
Display:

Display of min, max or hold values

Special display elements:
3 = Warning
triangle:

indicates a low battery

4 = Corr-arrow

:

indicates that correction factor is activated

5 =  Offset-arrow

:

indicates that zero point offset (offset) is activated

6 =  Min/Max/Hold

:

shows if a min., max. or hold value is displayed in
the secondary display

7 =  Alarm-arrow

:

Flashes if alarm is present

   

8 =  Logg-arrow

:

Shown if logger function is selected,
flashes if cyclic logger is running

1.5 Pushbuttons

key 1:

On/Off key

key 4:

Set/Menu

press (Menu) for 2 sec.: configuration will activated

keys 2, 5:

min/max when taking measurements:

press  shortly: min. or max. measuring value will be displayed
press for 2 sec.: the min. or max. value will be deleted

up/down for configuration:

to enter values or change settings

key 6:

Store/Quit

- Measurement: Hold current measuring value ('HLD' in display) or

operation of logger functions (p.r.t. Chapter 5)

- Menu: Acknowledge setting, return to measuring

key 3:

When taking measurements

: no function

During configuration

: Selection of menu-parameter

!

Summary of Contents for G 1700

Page 1: ...H49 0 02 6C 04 Operating Manual Precision Thermometer GMH 3750 from Version 1 0 For Pt100 4 Wire Temperature Probes...

Page 2: ...c conditions than those stated under Specification 2 Transporting the device from a cold to a warm environment condensation may result in a failure of the function In such a case make sure the device...

Page 3: ...terface or as analogue output This choice has to be done in the configuration menu 1 3 Connections 1 Output Operation as interface Connect to optically isolated interface adapter accessory GRS 3100 or...

Page 4: ...probe or measuring device oFF Scale correction factor inactive 0 000 4 4 1 120 Power off delay in minutes Device will be automatically switched off as soon as this time has elapsed if no key is press...

Page 5: ...res which deviate very much from the ambient temperature measuring errors often occur if the heat loss caused by the probe is not considered When measuring fluids therefore the probe should be emerged...

Page 6: ...arrow in the display shows an active zero displacement 4 4 Scale Correction Scale The scale of the measuring can be influenced by this setting factor is in displayed temperature C measured temperature...

Page 7: ...ead meas type in display 260 Logger read data manual logger 4 5 2 Analogue Output Scaling with DAC 0 and DAC 1 With the DAC 0 and DAC 1 values the output can be rapidly scaled to Your efforts Keep in...

Page 8: ...g the configuration an additional menu REAd LoGG is displayed see below or by means of the interface and a PC with GSOFT3050 software Max number of measurings 99 A measuring contains current measuring...

Page 9: ...e logger memory is full the display will show The recording automatically will be stopped Stopping the recording manually By pressing Store key 6 the recording can be stopped manually Then the followi...

Page 10: ...uring range specified temperature too high Wrong probe connected Check probe Err 1 sensor cable defective replace Value below display range Check Is the value below the measuring range specified tempe...

Page 11: ...lay and interface Real time clock Integrated clock with date and year Logger 2 Functions individual value logger Func Stor and cyclic logger Func CYCL Memory Stor 99 data sets CYCL 16384 data sets Cyc...

Page 12: ...H49 0 02 6C 04 Operating Manual GMH 3750 Page 12...

Reviews: