59
6.4 Communication Error Messages
Message Description
prtyXerror
This indicates that the parity of a received character did not
match the parity specified in the COMM PROTOCOL. This
could also result if the BAUD RATE is incorrect. The "X" in the
error message will be over-written with the associated serial port
number. 1= COM PORT 2 = PRINTER
ovrnXerror
This indicates an overrun error where an additional character was
received while the receive buffer of the Agrifeed 5500 was full
and, thus, the extra received character will be lost. The "X" in the
error message will be over-written with the associated serial port
number. 1= COM PORT 2 = PRINTER
frmgXerror
This indicates that the stop bit of a received character did not
occur when it was expected. This could be the result of an
incorrect
BAUD RATE
or
PROTOCOL
The "X" in the error
message will be over-written with the associated serial port
number. 1= COM PORT 2 = PRINTER
portXerror
This indicates that the Agrifeed 5500 did not check its receive
data register in time, thus missing a character. If this error should
occur, please notify your GSE dealer or the factory. To prevent
the problem, try reducing the BAUD RATE. The "X" in the error
message will be over-written with the associated serial port
number. 1= COM PORT 2 = PRINTER
6.4.1 Data
Transmissions
If a data transmission of any weight-related numeric data such as Gross,
Net or Tare, is sent to the remote as dashes, an overload or under load
(negative overload) condition was in effect. Remove the cause of the
overload (or under load) and repeat the transmission.
6.4.2 Displayed
Weight
If an overload or underload occurs due to an electrical overstress (EOS)
normally due to lightning or ESD discharge, press
[CLR]
. The message
"wait 1"
will appear for about 1 second. The A/D converter will then be
reset and the system should again be functional. If not, power down for a
few seconds. If the indicator still does not work properly after power-up,
check the load cell or platform wiring.