11. Fault finding

Observe the Oxiperm Pro installation and operating instructions.

Fault

Cause

Remedy

1. Display frozen.

a) Cable break.

Rectify

 

the cable break.

b) Calibration fault.

Repeat the calibration.

c) Incorrect analysis values used for calibration.

Repeat the calibration with correct values.

2. Measured value 

fluctuating shortly after 
calibration.

a) Measuring cell not allowed to run in correctly 

before calibration.

Allow the cell to run in for one hour before 
calibration.

3. Severely fluctuating 

measured value.

a) Cable and/or connectors corroded.

Replace the cable(s)/connector(s), and replace 
the electrode, if necessary.

4. Measured value too high 

or too low.

a) Air bubbles in electrolyte.

Unscrew the electrolyte chamber, and remove 
any air bubbles by carefully tapping the 
electrolyte chamber.

b) Not enough electrolyte in electrolyte chamber.

Unscrew the electrolyte chamber, fill up with 
electrolyte and recalibrate.

c) Severe change in temperature of sample water.

Recalibrate.

d) Deposits on electrode finger.

Unscrew the electrolyte chamber. Rinse the 
electrode finger, dry it with a dry paper towel, 
and carefully clean just the tip of the electrode 
finger with the special emery paper, then 
recalibrate.

e) Air bubbles in front of measurement opening 

(sample water side).

Check the water supply line.

Measuring module

5. Either no or not enough 

sample water flowing 
through measuring 
module.

a) Main water line closed off.

Open

 

the main water line.

b) No sample water present.

Check the sample water extraction point and 
supply line.

c) Isolating valve on measuring module input and/or 

output closed.

Open the isolating valve.

d) Circulating pump not functioning.

Switch on the circulating pump.

e) Overheating protection of circulating pump has 

been triggered.

Allow the circulating pump to cool off, then 
switch it back on.

f)

Circulating pump defective.

Replace the circulating pump.

6. Not enough sample 

water flowing through 
flow fitting / AQC-D6 
measuring cell.

a) Circulating pump not running. See above.

See above.

b) Main current of the measuring module not 

sufficiently throttled.

Throttle the main current more heavily using the 
throttle valve until the sample water flow rate is 
OK.

Summary of Contents for AQC-D6

Page 1: ...AQC D6 ClO2 measuring cell measuring module Installation and operating instructions GRUNDFOS INSTRUCTIONS...

Page 2: ...ation 8 9 Commissioning 8 9 1 Water connections 8 9 2 Electrical connections 9 9 3 Preparing the measuring system 10 9 4 Starting up the measuring system 11 9 5 Starting up the measuring module 11 9 6...

Page 3: ...s are used for measuring the concentration of chlorine dioxide in water within the scope of the potential applications described in this manual 5 Safety 5 1 Obligations of the owner operations manager...

Page 4: ...esponse time T90 30 seconds Temperature drift Temperature compensated measuring signal Permissible process water temperature 5 to 70 C Sample water flow rate Minimum 30 l h Maximum permitted pressure...

Page 5: ...3 Dimensional sketch drilling diagram Fig 3 AQC D6 Fig 4 Measuring module All dimensions in mm TM06 8111 4616 300 260 360 395 58 10 TM06 8112 4616 Height Width Depth 650 550 180 550 500 450 10 5 DN20...

Page 6: ...er The measuring cell is controlled with a potentiostat integrated in the measuring amplifier A precisely defined potential of the measuring electrode is retained through a reference system This resul...

Page 7: ...stream of sample water through the bypass It is designed specifically for compensating the internal pressure drops The extraction and feed line must therefore be kept as short as possible Sample wate...

Page 8: ...nnect hose 6 8 mm to a suitable drain Feed the fluid back to the pipework 9 1 2 Measuring module water connections The standard scope of delivery includes one PP and one PVC insertion section each for...

Page 9: ...Electrical connections may only be set up by authorised personnel Warning Switch off the power supply before connecting the power supply cable and the relay contacts Observe the local safety regulati...

Page 10: ...pos 2 Loosen the PDVF ring with 1 thread pos 1 Fit the measuring cell pos 3 in the flow fitting pos 2 1 Remove the protection cap 2 Slide the measuring cell into the flow fitting until the liquid flow...

Page 11: ...tons on the measuring amplifier in the Setup menu to select the Measuring cell line and press OK to access the corresponding menu 2 Use the Up and Down buttons to select the desired AQC D6 measuring c...

Page 12: ...calibration the regulators are switched off and the actuators closed to prevent overdosing 10 Operation 10 1 Switching on 10 2 Operation 10 3 Interruptions 10 3 1 Storage and handling when not in use...

Page 13: ...d Deposits on electrode finger Unscrew the electrolyte chamber Rinse the electrode finger dry it with a dry paper towel and carefully clean just the tip of the electrode finger with the special emery...

Page 14: ...5 Filling up with electrolyte 1 Screw the electrolyte chamber onto the measuring cell so that the socket of the electrolyte flange just fits into the gap approximately 5 mm 2 Fill electrolyte right up...

Page 15: ...namento delle legislazioni degli Stati membri UE NL EU conformiteitsverklaring Wij Grundfos verklaren geheel onder eigen verantwoordelijkheid dat de producten AQC D11 AQC D12 AQC D13 AQC D6 waarop de...

Page 16: ...16...

Page 17: ...ia S r l Via Gran Sasso 4 I 20060 Truccazzano Milano Tel 39 02 95838112 Telefax 39 02 95309290 95838461 Japan GRUNDFOS Pumps K K 1 2 3 Shin Miyakoda Kita ku Hamamatsu 431 2103 Japan Phone 81 53 428 47...

Page 18: ...splayed in this material including but not limited to Grundfos the Grundfos logo and be think innovate are registered trademarks owned by The Grundfos Group All rights reserved 2019 Grundfos Holding A...

Reviews: