background image

Euro-System Separator  

GENO

 

DK 2-Mini

 

 

 

 

 

 

 

Order no. 085 133 944-inter  Edited by: fger/mrie G:\BA-133944-INTER_085_SYSTEMTR-DK_2-MINI.DOCX 

4 | Technical specifications 

Table C-1: Technical specifications 

 

Euro system separator GENO-DK 2-Mini 

 

 

 

Connection data 

 

 

Nominal connection diameter 

[DN] 

15 

Threaded connection (male thread) 

[R] 

½“ (red. ¾“) 

Performance data 

 

 

Nominal pressure  

[bar] 

PN 10 

Min. flow pressure 

[bar] 

Max. flow 

[m³/h] 

3.5 

KV value  (

p = 1 bar) 

[m³/h] 

Dimensions and weights 

 

 

A  Installation length with screw connections 

[mm] 

130 

B  Installation length without screw connections 

[mm] 

83 

C  Drain connection HT pipe 

[mm] 

40 

D  Height above centre of pipe connection 

[mm] 

47 

E  Height below centre of pipe connection 

[mm] 

80 

F  Total height 

[mm] 

127 

G  Total width  

[mm] 

54 

H  Distance to wall 

[mm] 

35 

H  Space required above centre of pipe connection (e. g. for maintenance) 

[mm] 

200 

L  Clearance for the drain connection has to be provided in accordance  
    with the planned running of the pipes 

[mm] 

Observe installation requirements! 

Empty weight 

[kg] 

0.9 

Operating weight 

[kg] 

1.2 

Consumption data 

 

 

Dripping water volume per opening and closing process, approx. 

[ml] 

10 

Max. waste water volume at drain connection 

[m³/h] 

1.5 

Test certificate/Certification mark 

 

 

DVGW registration number 

 

NW-6305BR0345 

Ambient data 

 

 

Max. water temperature 

[°C] 

60 

Max. ambient temperature

 

[°C] 

60 

Order no. 

 

133 100 

 

 

Fig. C-2: Dimensional drawing  
of Euro system separator GENO-DK 2-Mini

 

Fig. C-3: Pressure loss curve of  
Euro system separator GENO-DK 2-Mini

 

 

5 | Scope of supply 

 

  Euro system separator GENO DK 2-Mini  

  Strainer for coarse dirt on inlet side  

  2 screw connections R ½“  

  2 reduction pieces R ¾“ male thread – R ½“ female thread  

  Flat seals 

  Operation manual

 

 

Summary of Contents for GENO DK 2-Mini

Page 1: ...Operation manual Euro system separator GENO DK 2 Mini Euro system separator GENO DK 2 mini Nominal diameter DN 15 Edition July 2017 Order no 085 133 944 inter...

Page 2: ......

Page 3: ...chments Warranty card Publisher s information All rights reserved Copyright by Gr nbeck Wasseraufbereitung GmbH Printed in Germany Effective with the date of edition indicated on the cover sheet We re...

Page 4: ...9 9074 41 120 Email service gruenbeck de We can connect you with the appropriate expert more quickly if you provide the required system data In order to have the required data handy at all times pleas...

Page 5: ...an installation directory of a water company as per 12 2 AVBWasserV German Ordinance on General Conditions for the Supply of Water 4 Shipping and storage Attention The systems and devices may be damag...

Page 6: ...cluding water originating from a drinking water installation that may exhibit modifications with regard to taste smell colour or temperature heating up or cooling down Examples Coffee heated drinking...

Page 7: ...hapter B classification of liquid categories Attention System separators are safety devices They protect the drinking water from modified drinking water They can only work reliably provided inspection...

Page 8: ...m 54 H Distance to wall mm 35 H Space required above centre of pipe connection e g for maintenance mm 200 L Clearance for the drain connection has to be provided in accordance with the planned running...

Page 9: ...xperts Attention Dirt and corrosion particles might cause malfunctions or damage the system separator Prior to start up flush the feed pipe at maximum performance 2 Water installation Binding rules Th...

Page 10: ...re that the discharge valve item 5 is closed 2 Completely deaerate the system separator by opening the bleeding screws item 3 3 Perform a visual check Make sure that no water leaks from the device any...

Page 11: ...steps Test step 1 Close shut off valve on outlet side Permanent discharge Non return valve on inlet side or dis charge valve leaky Remove and check Put system separator into operation again No drainin...

Page 12: ...service authorised service company Note A maintenance contract ensures that all the required maintenance work will be performed in due time 2 Inspection Check for tightness visual check If no water is...

Page 13: ...wearing parts Note Although these are wearing parts we grant a limited warranty period of 6 months Fig F 1 Spare parts for Euro system separator GENO DK 2 Mini Item Order no Designation PU Remarks 1...

Page 14: ...INTER_085_SYSTEMTR DK_2 MINI DOCX 12 Operation log Customer Name Address Euro system separator GENO DK please check appropriate box Serial number Installed by 1 1 1 2 DN 50 DN 65 DN 80 DN 100 DN 150...

Page 15: ...__________________________________________________________________ __________________________________________________________________ _________________________________________________________________...

Page 16: ...__________________________________________________________________ __________________________________________________________________ _________________________________________________________________...

Page 17: ...__________________________________________________________________ __________________________________________________________________ _________________________________________________________________...