Page 40
OCF 5.0 Open Channel Flow Monitor
Optional Sensor PZ32T
Maximum Range:
32 ft. (10 m)
Deadband (blanking):
Programmable, minimum 12” (305
mm)
Beam Angle:
8° at 3 DB
Temperature Compensation:
Automatic, continuous
Operating Frequency:
42 KHz
Exposed Materials:
PVC and Teflon
Operating Temperature:
- 40° to 150°F (-40° to 65°C)
Operating Pressure:
20 psi (1.35 Bar) maximum
Mounting:
¾” NPT (PVC isolation coupling
supplied)
Sensor Cable:
RG62AU coaxial, 25 ft. (7.6 m)
standard
Hazardous Rating:
with optional Intrinsic Safety Barrier:
CSA, Class I,II,III, Div. I,II, Groups
C,D,E,F,G
Note: Max Range reduced to 25 ft
(7.6 m) with ISB option.
Optional Sensor PZ32TE
Maximum Range:
32 ft. (10 m)
Deadband (blanking):
Programmable, minimum 12” (305
mm)
Beam Angle:
8° at 3 DB
Temperature Compensation:
Automatic, continuous
Operating Frequency:
42 KHz
Exposed Materials:
Teflon
Operating Temperature:
-40° to 170°F (-40° to 76°C)
Operating Pressure:
20 psi (1.35 Bar) maximum
Mounting:
¾” NPT (Teflon isolation coupling
supplied)
Sensor Cable:
RG62AU coaxial, 25 ft. (7.6 m)
standard
Hazardous Rating:
with optional Intrinsic Safety Barrier:
CSA, Class I,II,III, Div. I,II, Groups
C,D,E,F,G,
Note: Max Range reduced to 25 ft
(7.6 m) with ISB option.
Summary of Contents for OCF 5.0
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 OCF 5 0 Open Channel Flow Monitor CONNECTIONS ...
Page 7: ...Page 7 OCF 5 0 Open Channel Flow Monitor CALIBRATION MENU ...
Page 8: ...Page 8 OCF 5 0 Open Channel Flow Monitor ICONS ...
Page 14: ...Page 14 OCF 5 0 Open Channel Flow Monitor CHANNEL SETUP ...
Page 20: ...Page 20 OCF 5 0 Open Channel Flow Monitor SENSOR MOUNTING METHODS ...
Page 33: ...Page 33 OCF 5 0 Open Channel Flow Monitor EXTENDED SENSOR CABLE INSTALLATION IN MANHOLE ...
Page 35: ...Page 35 OCF 5 0 Open Channel Flow Monitor ...
Page 41: ......