Page 27
OCF 5.0 Open Channel Flow Monitor
C
- sensor cable extended and junction not
insulated
- Use metal Junction Box
C
- through enclosure
- use metal enclosure
C
- through 4-20mA output cable
- use shielded twisted pair (shielded to AC ground)
- use grounded metal conduit
C
- wiring or installation close to variable
speed drive or inverter
- follow V.S.D. manufacturer’s instructions for Drive
grounding, wiring and shielding
Unit Receives No Return Echo Due To:
C,F,E
- foam on liquid surface
- use stilling well (open channel flow)
B
- target beyond
Max Range
- recalibrate
F
- sensor damaged
- remove sensor from mounting and aim at a flat,
stable target to test
F
- sensor misalignment
- check with a level
Wiring Problems Due To Sensor Cable:
A,C,F,
- open circuit
- check connections/continuity (8850 to 12700 ohms
max./-30°C to +70°C )
B,F
- short circuit
- check connections/continuity (8850 ohms min.)
F
- too long (max 500 ft., 150 m)
C
- bundled/run in conduit with power cable
C
- sensor ground shorted to conduit/enclosure
- insulate
A
- extended with wrong type of wire
- use RG62A/U coaxial only
C
- close to high voltage/large motors
C
- AC chassis/ground missing on instrument
power connections
Non-Linearity Due To:
H
- vapour
- dissipate fumes, Calibration in-situ
H
- zero not set accurately
- see “Zero Positioning of Sensor”
H
- wrong flume, or K&n selected (
FLOW
mode)
- select correct flume
- temperature measurement inaccuracy
- install sensor sunscreen
- mount sensor closer to flow
Summary of Contents for OCF 5.0
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 OCF 5 0 Open Channel Flow Monitor CONNECTIONS ...
Page 7: ...Page 7 OCF 5 0 Open Channel Flow Monitor CALIBRATION MENU ...
Page 8: ...Page 8 OCF 5 0 Open Channel Flow Monitor ICONS ...
Page 14: ...Page 14 OCF 5 0 Open Channel Flow Monitor CHANNEL SETUP ...
Page 20: ...Page 20 OCF 5 0 Open Channel Flow Monitor SENSOR MOUNTING METHODS ...
Page 33: ...Page 33 OCF 5 0 Open Channel Flow Monitor EXTENDED SENSOR CABLE INSTALLATION IN MANHOLE ...
Page 35: ...Page 35 OCF 5 0 Open Channel Flow Monitor ...
Page 41: ......