Page 11
MantaRay Portable Area-Velocity Flow Meter
CALIBRATION
Press
to
Calibration
and
to enter. Use
or
to position
before
each menu item and
to enter. When settings are completed press
to store and
return to the Calibration menu.
20mA Flo
[5V Flo]
[OFF]
Press
and enter the flow rate value for 20mA.
Note: Analogue output can be selected as 4-20mA or 0-5V in Special Functions.
4mA Flo
[0V Flo]
[OFF]
Press
and enter the flow rate value for 4mA.
Lvl Offset
Optional for QZ02L sensor (use for mud or silt conditions).
Press
and enter an offset to level measurement. Set to 0.00
when sensor mounted on floor of channel. When sensor is
mounted above the floor of the channel enter the distance
between channel floor and bottom of sensor. Maximum offset
is ± 36” (914 mm).
Note: 4mA is not affected by
Lvl Offset
settings. 4mA
is the bottom of the channel or pipe.
Min Vel
Press
and enter a minimum velocity cutoff. Forward and
reverse velocities less than
Min Vel
will be forced to zero.
Damping
Increase damping to stabilize readings under turbulent flow
readings or to reject spurious level readings. Decrease for
faster response to changes in flow.
Press
from the
Calibration
display to return to Menu Selections.
--Calibration-------
}
20mA Flo
10.000ft /s
³
4mA Flo
0.000ft /s
³
Min Level
0.000in
Lvl Offset
0.000in
Min Vel
0.000ft/s
Damping
10%
SENSOR
END VIEW
SIDE VIEW
TOP VIEW
SENSOR
U-SHAPE STAINLESS STEEL
18 GA. RECOMMENDED
HEIGHT AS REQUIRED
USE THE MB-QZ
STAINLESS STEEL
MOUNTING BRACKET
(SUPPLIED)
SENSOR ELEVATED
ABOVE FLOOR OF
PIPE OR CHANNEL
LEVEL
SENSOR