ERROR/WARNING MESSAGES
Page 18
LIT25 Level Indicating Transmitter
Manual Series 3.5.1
Instrument has detected Sensor connections/cable open.
Instrument has detected Sensor connections/cable short.
No valid echoes for 10 seconds.
No valid echoes for 20 seconds. See Troubleshooting (F).
Illegal Span. Distance between 20 = and 4 = must be
greater than 2" (5 cm)
Indicates that the target is above the
calibrated maximum level (too close to
the Sensor).
Indicates target is further than the calibrated Maximum Range.
Indicates that the LIT25 has experienced electrical interference strong enough to corrupt
the memory. The LIT25 must be reset and recalibrated.
: Reset will clear all memory. LIT25 will need recalibration after this
procedure: Press and hold
Reset Procedure 1
and
until the LIT25 displays - - - -.
Ç
È
X X
Always displayed at power-up (x.x indicates Software version).
Otherwise indicates instrument has reset automatically. See Troubleshooting (C,D,E).