SENSOR MOUNTING LOCATION
SENSOR MOUNTING
Each sensor is equipped with a ¾ inch “isolation coupling” which MUST be used in your installation. A
threaded nipple or length of conduit may be used to position the sensor at the desired height.
The sensor should be hand-tightened (like a light bulb) by turning the sensor stem only. DO NOT use a
wrench and do not over tighten. DO NOT clamp the sensor below the isolation coupling.
The standard PZ32T Sensor must be mounted 12" / 30.48 cm or more above the maximum liquid level.
(Optional PZ12 sensor must be 8" / 20.3 cm above maximum level and optional PZ52T sensor must be
16" / 406 mm above maximum level.)
Page 15
LIT25 Level Indicating Transmitter
Manual Series 3.5.1
1 FT HORIZONTAL
FROM SIDEWALL FOR
EVERY 10 FT VERTICAL
NOTE: 2 FT FOR EVERY
10 FT DEPTH ON
APPLICATIONS WITH
ROUGH SIDEWALLS,
LADDERS,
REINFORCING RINGS
ETC.
PREFERRED SENSOR
LOCATION
6" SCHEDULE 40
PIPE IF EXTENSION
REQUIRED
SENSOR FACE MUST
BE HORIZONTAL
±1°
MAX. LIQUID LEVEL