SPECIAL FUNCTIONS
DFM-IV ver
shows software version installed
Tag
enter instrument Tag Number (0-9999)
Date
(with Optional Data Logger) use the
È
or
Ç
keys to change date as required
Time
(with Optional Data Logger) use the
È
or
Ç
keys to change time as required
Reset Tot? Yes
Select
Yes
to reset totalizer
Defaults? Yes
Press
Æ
3 times to select
Yes
. Store to erase all user settings and return
instrument to factory default settings.
Simul 0.00%
exercises 4-20mA output, digital display and control relays. Simplifies
calibration of remote devices on the 4-20mA loop and checks set
point/operation of Relays calibrated in Flow mode. Go
È
for
100%
.
4mA ADJ
20mA ADJ
Use to fine tune the 4mA and 20mA calibration of the 4-20mA output.
Adjustment range is +1mA to –1mA in 0.002mA steps
Important: The 4-20mA output will be forced to 4mA and 20mA respectively
during these adjustments.
Master or Slave
Master or Slave selection for synchronization of multiple DFM-IV flow meters
(see Manual section SYNCHRONIZATION).
New Password
position cursor under digits and set new number between
00
and
99
Com 24 48 96 192
(Optional) set the baud rate of the DFM-IV RS232 output when equipped with
an optional data logger.
DFM-IV Doppler Flow Meter
Manual Series A.6
Page 10
SPECIAL FUNCTION
DFM-IV ver 2.67
TAG
09
Date Jan 01/2000
Time
11:50:57
Reset Tot?
Yes
Defaults?
Yes
20mA ADJ 20.000
Master
Slave
New Password: 00
Com 24 48 9
192
6
Store?
Yes
*** STORING ***
Simul
0.00 %
4mA ADJ
4.000