RS232C SERIAL OUTPUT (with optional Data Logger)
Format: 8 Bits, 1 Stop Bit, No Parity. (Baud rate may be changed under the
SPECIAL FUNCTIONS
menu. Factory default is 19200 baud).
Output connector is DB9-Female. Use the cable supplied for connection to a PC Computer. Use the
cable and optional Null Modem NM-DB9M/DB25M for connection to a modem.
Page 37
DFM-IV Doppler Flow Meter
Manual Series A.6
Greyline
Instrument
RS232
20' (6 m) cable
included
Detail
INCLUDED
RX
RS 232
PU GND TX
(GR
EEN)
(W
HT
)
(BL
K
)
(RE
D
)
Wired as DCE
(Data Communication Equipment)
PIN 2 - RX
PIN 3 - TX
PIN 5 - GND
SHIELD - CASE
(PU = PULL-UP)
PIN 6 - DSR (PU)
PIN 8 - CTS (PU)
5
4
3
2
1
9
8
7
6
Detail
DB9
F
DB25
F
Modem
Modem
Computer
DB9M
DB25
M
Public Telephone Lines
NULL MODEM
Option
NM-DB9M/DB25M