DFM-IV Doppler Flow Meter
Manual Series A.6
Relays 1 2 3 (Relay Status Display)
Displays state of each Relay.
1
2 3 4 5
- Reverse character indicates that the Relay is On (energized).
24 HR LOG
DFM-IV Flow Meters with optional Data Logger will display this menu. Refer to OPTIONS section.
PASSWORD
The password (a number from
00
to
99
) prevents unauthorized access to the
CALIBRATION
menu.
From
RUN
press
Æ
to get to
PASSWORD
.
Press
Æ
to place the cursor under the digit and
È
or
Ç
to enter your password.
Factory default Password is
00
.
A new password can be stored by going to
SPECIAL FUNCTIONS
and
È
to
New Password
.
UNITS /MODE
Use
Æ
to get to
UNITS / MODE
, then
È
to get to Linear Units. Use
Æ
to move the cursor under the
required units.
Use
È
to get to Volume Units and then select units of Time and
Flow
or
Velocity
.
Flow
mode displays flow rate in engineering units (e.g. gpm, litres/sec etc.)
Velocity
mode displays flow velocity in units/time (e.g. ft/sec or m/sec)
When all units have been selected go to
Store
?
then
Æ
to
Yes
and
È
or
Ç
to
CALIBRATION
.
Page 7
PASSWORD: 00
PASSWORD:
0
0
PASSWORD:
0
1
UNITS / MODE
PASSWORD
Flow
Velocity
ft
in
m
cm
Ft3 USG USMG IG>
IMG
m3
L
brl
s
min
Hr
d
Store?
Yes
*** STORING ***
CALIBRATION