Calibration 

In order to compensate for ageing of the sensor and air pressure fluctuations due 
to weather changes, the sensor has to be calibrated at regular intervals. The 
sensor will be calibrated to the atmospheric oxygen concentration of 20.95%. 
We recommend to calibrate directly before starting the measuring process.  

Calibration:

.

The sensor has to be subjected to air (make sure that rooms are 

thoroughly aired). 

Press the 

 key for 2 sec till 

 appears. 

The calibration will be automatically completed as soon as the measuring value is 
stable (takes a few seconds).  Afterwards the rating of the sensor state will be 
shown for a short time. 

If an error message (CE.3, CE.4, CE.6) is displayed, the sensor signal is 
invalid. Press any key to restart, the previous calibration will be restored. 
The display is blinking afterwards until a valid calibration is performed. 

10  Sensor rating 

The rating of the sensor state is calculated and saved after successful calibration.  

Watch sensor rating:

 Press shortly the 

 

key, the display shows for a short 

time i.e. 

 = 100%. 

The rating is displayed in 10 percent steps: 100% = optimal sensor status. Lower values are 
indicating that the sensor life time will soon be reached. (50% doesn’t mean 50% of the life 
time, but 50% of the reference signal!). Erroneous P.AbS entries may cause to low ratings

11  Errors and system messages 

no  

display 

The battery is used up and has to be exchanged, or the device is defective 

blinking 

display 

The last calibration was not successful. A new calibration has to be performed! 

 

The battery is used up, for a short period further measurements are possible 

 

The battery is used up and has to be replaced, measurements are no more 
possible  

 

Calibration error: sensor voltage is too low (sensor used up or wrong chosen air 
pressure) 

 

Calibration error: sensor voltage is too high (wrong chosen air pressure or 
sensor defective (membrane crack, etc.) 

 

Calibration error: signal not stable 

 

Measuring error: range has been exceeded 

 

Measuring error: measuring value has fallen below permitted range 

 

System error: the device is defective or considerably out of the allowed ambient 
temperature range 

 

Value could not be calculated (e.g. input of MOD-calculation is error) 

 
 

 

Summary of Contents for GOX 100T

Page 1: ...k GmbH Standort Greisinger Hans Sachs Stra e 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Oxygen Measuring Device for diving gas mixtures WEEE Reg Nr DE 93889386 as of V...

Page 2: ...s will be detected at calibration and the sensor element has to be replaced before the next measurement By means of the integrated hold function the MOD Display max operating depth and a specialized s...

Page 3: ...y to avoid re starting Operator safety may be a risk if there is visible damage to the device or the device is not working as specified the device has been stored under unsuitable conditions for a lon...

Page 4: ...has to be replaced measuring is valid On Off switch Press shortly change between actual value oxygen hold value oxygen hold value MOD in meters n or feet F Press shortly short display of the sensor ra...

Page 5: ...out at other pressure conditions than the calibration The actual pressure has to be entered in the configuration for measurement and calibration The sensor s temperature dependency is compensated inte...

Page 6: ...Press shortly the key the display shows for a short time i e 100 The rating is displayed in 10 percent steps 100 optimal sensor status Lower values are indicating that the sensor life time will soon b...

Page 7: ...e 1 to 120 min OFF auto power off is deactivated continuous operation Abs air pressure factory setting 1013 mbar 500 1200 Input of absolute pressure of measured gas The absolute pressure is depending...

Page 8: ...ed strain max 0 25 bar over under pressure Display approx 13 mm high 3 digit LCD Pushbuttons 3 keys for ON OFF min max value display calibration Nominal temp 25 C Ambient temp 0 to 45 C sensor 20 to 5...

Reviews: