Oxygen measuring 

Switch on instument, calibrate (refer to chapter 9 Calibration) 
-> the

 current oxygen concentration [%] 

will be displayed 

hold sensor to valve like shown (good 
contact necessary, otherwise there 
could be measured to low values) 

Open valve a little, until You can hear 
gas flow.  
Attention: Do not open to far! 

When measurement display is stable, 

press 

 -key: 

 

hld

 (Hold = measurement is held) and 

oxygen concentration [%]

 are displayed alternating 

pressing 

 again: 

 

The Maximum Operating Depth (

MOD

) is displayed alternating to 

hld 

(

n

 is the symbol for display in meters, 

F

 for feet) 

pressing 

 again: return to current oxygen concentration display 

 

The oxygen measurement 

depends on air pressure

. The current pressure can be entered 

during configuration.  

If you 

measure at atmosphere

 and the calibration took place at the same conditions, the 

pressure value has not to be adjusted necessarily, because the same pressure is used for 
calibration and measuring.  

(If the measurement is carried out at

 other pressure conditions

 than the calibration: The 

actual pressure has to be entered in the configuration for  measurement and calibration. ) 
The sensor’s

 temperature dependency

 is compensated internally. However for best results 

try to keep the same temperature both during calibration and measuring. 

Temperature differences between sensor and gas

 may cause additional errors! 

Therefore wait an appropriate period of time until the sensor has adjusted its temperature to 
the gas that is measured. A suitable flow of the gas around the sensor increases the 
adjustment significantly.  
Try to avoid warming of the sensor by touching it during measurement or calibration. 

When 

measuring bottled gas

, consider that the gas coming out is cooler than the ambient 

because of the pressure loss! Strong air flows can produce an over pressure at the sensor 

– 

possible source of measuring/calibration errors! 

The nominal life time can be shortened by: 

Wrong storage / operation temperature 

Permanent use with dry gases (compressed gas). 
It helps to put the sensor to normal-

humid ambient air in measuring breaks (“flush” 

system with fresh air). 

The 

optimum operation position

 is: with the sensor inlet pointing downwards 

Summary of Contents for GOX 100T

Page 1: ...k GmbH Standort Greisinger Hans Sachs Stra e 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Oxygen Measuring Device for diving gas mixtures WEEE Reg Nr DE 93889386 as of V...

Page 2: ...s will be detected at calibration and the sensor element has to be replaced before the next measurement By means of the integrated hold function the MOD Display max operating depth and a specialized s...

Page 3: ...y to avoid re starting Operator safety may be a risk if there is visible damage to the device or the device is not working as specified the device has been stored under unsuitable conditions for a lon...

Page 4: ...has to be replaced measuring is valid On Off switch Press shortly change between actual value oxygen hold value oxygen hold value MOD in meters n or feet F Press shortly short display of the sensor ra...

Page 5: ...out at other pressure conditions than the calibration The actual pressure has to be entered in the configuration for measurement and calibration The sensor s temperature dependency is compensated inte...

Page 6: ...Press shortly the key the display shows for a short time i e 100 The rating is displayed in 10 percent steps 100 optimal sensor status Lower values are indicating that the sensor life time will soon b...

Page 7: ...e 1 to 120 min OFF auto power off is deactivated continuous operation Abs air pressure factory setting 1013 mbar 500 1200 Input of absolute pressure of measured gas The absolute pressure is depending...

Page 8: ...ed strain max 0 25 bar over under pressure Display approx 13 mm high 3 digit LCD Pushbuttons 3 keys for ON OFF min max value display calibration Nominal temp 25 C Ambient temp 0 to 45 C sensor 20 to 5...

Reviews: