Operating and maintenance advice 

  The battery has to be taken out, when the device is stored above 50°C. 
  Handle the device and the sensor with care and only use it according its 

specification. The connectors have to be saved from humidity and dirt.  

  Unplug the sensor not by pulling the cable, but the plug.  

Tip: It is recommended to take the battery out, when the device is stored for a 

longer period of time. 

 

Operation 

Display- and Operation elements 

 

1.  display of the oxygen concentration in % or of 

MOD in meters(n) or feet(F)  

2.  BAT-warning: battery is used up and has to be 

replaced (measuring is valid) 

 

 

On/Off switch

  

 

Press shortly: change between  

                    : 

actual value % oxygen 

: hold value % oxygen,

                      

: hold value MOD in meters(n)  

                                                                  or feet(F)

 

 

Press shortly: short display of the sensor rating (calculated after the last 

calibration) (i.e. “100.P“ = 100%) 

Press long:     start calibration (see also chapter 9 Calibration) 

First start  

If not already done: plug the sensor to the device. 
Screw flow-diverter on sensor, plug on transparent T-piece 

Switch the device on with the

 

 key. 

Start the calibration by pressing the

 

 key for 2 seconds, until 

 is 

displayed. After the calibration the device is ready for measuring.  
(see also chapter 9 Calibration) 

Display when switching-on the device 

 

Start and display test 

 

If switch-

off delay “P.oF” (power off) is active, this is signalled at the turn-on 

procedure.  

 

 

Summary of Contents for GOX 100T

Page 1: ...k GmbH Standort Greisinger Hans Sachs Stra e 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Oxygen Measuring Device for diving gas mixtures WEEE Reg Nr DE 93889386 as of V...

Page 2: ...s will be detected at calibration and the sensor element has to be replaced before the next measurement By means of the integrated hold function the MOD Display max operating depth and a specialized s...

Page 3: ...y to avoid re starting Operator safety may be a risk if there is visible damage to the device or the device is not working as specified the device has been stored under unsuitable conditions for a lon...

Page 4: ...has to be replaced measuring is valid On Off switch Press shortly change between actual value oxygen hold value oxygen hold value MOD in meters n or feet F Press shortly short display of the sensor ra...

Page 5: ...out at other pressure conditions than the calibration The actual pressure has to be entered in the configuration for measurement and calibration The sensor s temperature dependency is compensated inte...

Page 6: ...Press shortly the key the display shows for a short time i e 100 The rating is displayed in 10 percent steps 100 optimal sensor status Lower values are indicating that the sensor life time will soon b...

Page 7: ...e 1 to 120 min OFF auto power off is deactivated continuous operation Abs air pressure factory setting 1013 mbar 500 1200 Input of absolute pressure of measured gas The absolute pressure is depending...

Page 8: ...ed strain max 0 25 bar over under pressure Display approx 13 mm high 3 digit LCD Pushbuttons 3 keys for ON OFF min max value display calibration Nominal temp 25 C Ambient temp 0 to 45 C sensor 20 to 5...

Reviews: