GREISINGER GMH 5130 Operating Manual Download Page 9

H82.0.01.6C-02 

 Operating Manual GMH 5130 

page 9 of 14 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

7.2.1  Standard measuring (slow) 

 

Measuring rate 4 Hz, averaging and filter functions are active. 
Application: Measuring of slowly changing or static pressures, e.g. measuring of leakproofness, 
atmospheric pressure... 
Highest accuracy, high noise immunity, low power consumption. 

7.2.2  Peak detection (Peak detection) 

 

Measuring rate >1000 Hz, the value is displayed unfiltered. 
Application with logger function: Measuring of short pressure peaks or fast changing pressures with 
a resolution of <1 ms. The cyclic logger function records the arithmetic mean value, the highest and 
the lowest peak of the referring time interval. 

Attention: higher power consumption, measuring is sensitive to noise (EMI, ...). 

 

Measuring is sensitive to noise (EMI, ...)  
and higher power consumption. 

 

7.2.3  Fast filtered measuring (fast) 

 

Measuring rate >1000 Hz, but the value is filtered (higher noise immunity than P.dEt, small peaks 
will be filtered out), apart from that identical behavior like rAtE-P.dEt. 

7.3  Water depth / level measurement  

 

Water-proof sensors have to be used for water depth / level measurement. 

Select the unit [m] for meters water column (selection mH2O) in menu 

“Unit“. 10 m water column (=water 

depth) corresponds roughly 1 bar overpressure. 

Measurings can be made e.g. like described below: 

(for abs. pressure sensors SL has to be deactivated)  

  With an abs. pressure sensor: 

Press ‘Tara’ when sensor is at ambient air and then bring sensor to the 

depth to be measured. The display shows now the depth in [m]. 

  With a rel. pressure sensor: bring tube connection for lower press. in contact to ambient air by means of 

a tube (no water contact!) and bring the sensor with its open pressure connection for higher pressure to 
water depth to be measured (display and is compensated for pressure changes in ambient air). 

7.4  Sea-level correction for absolute pressure sensors  

The device displays the absolute pressure. This is not necessarily the same like the values given by weather 
stations! The weather stations are displaying the pressure at sea level. Usually the sensor is placed above 
sea level and therefore the pressure loss resulting from the actual level above sea level has to be 
considered, if the value at sea level (zero) should be measured! To correct the measuring display, activate 
the “Sea-Level-Function” (SL, p.r.t. chapter 0, setting is only possible, if an abs. pressure sensor is 
connected). Then enter the altitude above 

sea level of the sensor’s location in meters (Alti, p.r.t. chapter 6). 

When activated, the display shows the SL-arrow and the device displays the pressure value at sea level. 

7.5  Averaging function 

The averaging function concerns the display values (LCD and interface).  
  

The averaging integrates the measuring values during a selectable period of time and then calculates the 
average display value. It is independent from the selected kind of measuring (slow, fast, peak detect). 

As long as not enough values have been collected (i. e. for the selected averaging time) to calculate an 
average value

, the upper display shows “----“, the lower display a ‘countdown‘. 

Function of min/max-value memory during averaging: 

-   If averaging is activated and slow measuring is selected (rAtE-Slo), the min-/max-value memory refers to 

the average display value. 

-  If averaging is activated and fast measuring is selected (rAtE-FASt or P.dEt), the min-/max-value memory 

refers to the internal measured values (fast peaks can be detected). 

 

 

Summary of Contents for GMH 5130

Page 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de WEEE Reg Nr DE 93889386 as of version 1 0 GMH 5130 Operating Manual Handheld Pressure Meter water proof Please carefully read these instructions b...

Page 2: ...LEMENTS 5 4 2 PUSHBUTTONS 5 4 3 CONNECTIONS 5 4 4 POP UP CLIP 6 5 START OPERATION 7 6 CONFIGURATION 7 7 REMARKS TO SPECIAL FEATURES 8 7 1 DIFFERENT PRESSURE UNITS 8 7 2 DIFFERENT KINDS OF MEASURING 8...

Page 3: ...to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Prot...

Page 4: ...e usage of this device at potentially explosive areas increases danger of deflagration explosion or fire due to sparking 3 Product Specification 3 1 Scope of supply The scope of supply includes GMH 51...

Page 5: ...ase refer to chapter 7 1 7 corr arrow zero point or slope correction is active 4 2 Pushbuttons On off key backlight press shortly activate backlight or switch on instrument press longer switch off ins...

Page 6: ...to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attac...

Page 7: ...es the settings enter finishes the configuration and returns to standard measuring operation Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults...

Page 8: ...sor dep e g 5 00 5 00 mbar The offset of sensor will be displaced by this value to compensate for deviations of the probe or of the measuring device Slope adjustment of sensors oFF No slope adjustment...

Page 9: ...ing tube connection for lower press in contact to ambient air by means of a tube no water contact and bring the sensor with its open pressure connection for higher pressure to water depth to be measur...

Page 10: ...l net Universal multi channel software for real time recording and displaying of measuring data with real database In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...o G1 2 pressure terminal or plug plastic tube to a suitable adapter Connection to instrument Use cable MSD K51 11 Accuracy Check Adjustment Service You can send the device to the manufacturer for adju...

Page 12: ...h to be displayed Err 4 Display range underflow Check value below 19999 Tara to low Er 11 Value could not be calculated Choose different unit Calculation overflow happened Choose different unit Err 7...

Page 13: ...ment display illuminated white Correction function offset slope via menu if offset slope value not equal zero corr is displayed Additional functions min max hold function Averaging adjustable 1 120 se...

Page 14: ...H82 0 01 6C 02 Operating Manual GMH 5130 page 14 of 14 _____________________________________________________ _____________________________________________________________________________...