15 Specifications 

Display range 

maximal -19999...19999 digit, depends on used sensor 

Number of channels 

Suitable probes 

GMSD 

… - K51 and MSD … sensors with connection cable MSD-K51 

Available ranges/resolutions  

ranging from  -

1.999 … 2.500 mbar / 0.001 mbar  to  0.0 … 1000 bar / 1 bar 

Connections 

sensor 

7-pole bayonet connector 

output /  
ext. supply 

4-pole connector for serial interface and supply (USB Adapter USB 5100) 

Display units 

depends on meas. range selection (depends on connected sensor): 
mbar, bar, Pa, kPa, MPa, mmHg, inHg, PSI, mH

2

User unit 

freely adjustable 

Measuring frequency 

4 / s  or 1000 / s 

Accuracy 

± 0.1 % FS ± 1 digit 

Working conditions 

-25 ... 50 °C; 0 ... 95 %RH (non condensing) 

Storage conditions 

-25 ... 70 °C 

Display 

4 ½ digit 7-segment display, illuminated (white) 

Correction function 

offset / slope (via menu),  
if offset/slope value not equal zero: "corr" is displayed 

Additional functions 

min- / max- / hold- function 

Averaging 

 

adjustable

, 1 … 120 seconds 

Housing 

protection class  IP65 and IP67 

 

dimensions 
L*W*H [mm] 

160 * 86 * 37 incl. protective covering, 

approx. 250 g incl. batteries and protective covering 

Power supply 

 

2*AAA-batteries (included in delivery)  

current 
consumption 

2.0 mA (at Out = Off, equivalent to 500 h),  
illumination ~10mA (turns off automatically) 

Battery-change indicator 

automatically if battery exhausted "bAt", warning "bAt" flashing 

Auto-off-function 

Device will be automatically switched off if no key is pressed/no interface 
communication takes place for the time of the power-off delay.  
The power-off delay can be set to values between 1 and 120 min.; it can be 
completely deactivated. 

EMC 

The device corresponds to the essential protection ratings established in the 
Regulations of the Council for the Approximation of Legislation for the member 
countries regarding electromagnetic compatibility (2004/108/EG) 
Additional fault: <1%

 

 

 

Summary of Contents for GMH 5130

Page 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de WEEE Reg Nr DE 93889386 as of version 1 0 GMH 5130 Operating Manual Handheld Pressure Meter water proof Please carefully read these instructions b...

Page 2: ...LEMENTS 5 4 2 PUSHBUTTONS 5 4 3 CONNECTIONS 5 4 4 POP UP CLIP 6 5 START OPERATION 7 6 CONFIGURATION 7 7 REMARKS TO SPECIAL FEATURES 8 7 1 DIFFERENT PRESSURE UNITS 8 7 2 DIFFERENT KINDS OF MEASURING 8...

Page 3: ...to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Prot...

Page 4: ...e usage of this device at potentially explosive areas increases danger of deflagration explosion or fire due to sparking 3 Product Specification 3 1 Scope of supply The scope of supply includes GMH 51...

Page 5: ...ase refer to chapter 7 1 7 corr arrow zero point or slope correction is active 4 2 Pushbuttons On off key backlight press shortly activate backlight or switch on instrument press longer switch off ins...

Page 6: ...to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attac...

Page 7: ...es the settings enter finishes the configuration and returns to standard measuring operation Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults...

Page 8: ...sor dep e g 5 00 5 00 mbar The offset of sensor will be displaced by this value to compensate for deviations of the probe or of the measuring device Slope adjustment of sensors oFF No slope adjustment...

Page 9: ...ing tube connection for lower press in contact to ambient air by means of a tube no water contact and bring the sensor with its open pressure connection for higher pressure to water depth to be measur...

Page 10: ...l net Universal multi channel software for real time recording and displaying of measuring data with real database In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...o G1 2 pressure terminal or plug plastic tube to a suitable adapter Connection to instrument Use cable MSD K51 11 Accuracy Check Adjustment Service You can send the device to the manufacturer for adju...

Page 12: ...h to be displayed Err 4 Display range underflow Check value below 19999 Tara to low Er 11 Value could not be calculated Choose different unit Calculation overflow happened Choose different unit Err 7...

Page 13: ...ment display illuminated white Correction function offset slope via menu if offset slope value not equal zero corr is displayed Additional functions min max hold function Averaging adjustable 1 120 se...

Page 14: ...H82 0 01 6C 02 Operating Manual GMH 5130 page 14 of 14 _____________________________________________________ _____________________________________________________________________________...

Reviews: