 Operating Manual GMH 5130 

page 8 of 14 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

 

 

Averaging Filter 

 

1 … 120 

Averaging period in seconds  

oFF 

Averaging function is deactivated 

 

Auto Power-Off: Select power-off delay

 

 

1 ... 120 

Power-off delay in minutes.  
Device will be automatically switched off as soon as this time has 
elapsed if no key is pressed/no interface communication takes place.  

oFF 

Automatic power-off function deactivated (continuous operation) 

 

Background illumination 

 

oFF  

Illumination deactivated 

5 … 120 

Turn off illumination after 5 … 120s (factory settings: 5 s) 

on 

Illumination always on 

 

Set Output: Universal output adjustments  

 

 

oFF 

Interface and analog output off 

 

SEr:  

Serial interface activated

 

 

01, 11..91 

Base address for serial interface communication  
(only at Out = SEr)

 

 

 

Set Corr: Input adjustment 

 

 

 

 

 

 

 

 

 

 

 

 

Zero adjustment/offset of sensors 

 

oFF 

No zero adjustment of sensors 

Sensor dep.

 

e.g.-5.00..5.00 mbar

 

The offset of sensor will be displaced by this value to compensate for 
deviations of the probe or of the measuring device. 

 

Slope adjustment of sensors 

 

oFF 

No slope adjustment of sensor 

-

2.000 … 2.000 

The slope of sensor will be changed by this factor  [%] to compensate 
deviations of probe or measuring device 

 (**) 

Parameter can only be called if corresponding sensor is plugged. 

 

7  Remarks to Special Features  

7.1  Different Pressure Units 

Depending on the connected sensors different units can be selected via menu (Unit). The measuring range 
of the sensors may restrict the choice! 

User-Unit 

For pressure units, which are not covered by the ones printed on the display, there can be done a manual 
setting for a 

“user-defined” unit. 

Example: 

To enable a GMSD 2 BR - K51 ( -1000 

… 2000 mbar) to display the unit kg/cm

2

, following settings (bold) 

have to be entered: 

 

kg/cm

2

 

Torr 

atm 

at 

 

bASE user:  

bar 

mbar 

bar 

bar 

 

DP user:  

.0000 

.0000 

.0000 

.0000 

 

FACt user: 

1.0197 

.7433 

0.9869  1.0197   

7.2  Different kinds of measuring  

Three different kinds of pressure measuring are supported. Two of them (P.dEt and FASt) are working with 
high measuring frequency of more than 1000 measurings per second.  
 

 

Summary of Contents for GMH 5130

Page 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de WEEE Reg Nr DE 93889386 as of version 1 0 GMH 5130 Operating Manual Handheld Pressure Meter water proof Please carefully read these instructions b...

Page 2: ...LEMENTS 5 4 2 PUSHBUTTONS 5 4 3 CONNECTIONS 5 4 4 POP UP CLIP 6 5 START OPERATION 7 6 CONFIGURATION 7 7 REMARKS TO SPECIAL FEATURES 8 7 1 DIFFERENT PRESSURE UNITS 8 7 2 DIFFERENT KINDS OF MEASURING 8...

Page 3: ...to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Prot...

Page 4: ...e usage of this device at potentially explosive areas increases danger of deflagration explosion or fire due to sparking 3 Product Specification 3 1 Scope of supply The scope of supply includes GMH 51...

Page 5: ...ase refer to chapter 7 1 7 corr arrow zero point or slope correction is active 4 2 Pushbuttons On off key backlight press shortly activate backlight or switch on instrument press longer switch off ins...

Page 6: ...to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attac...

Page 7: ...es the settings enter finishes the configuration and returns to standard measuring operation Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults...

Page 8: ...sor dep e g 5 00 5 00 mbar The offset of sensor will be displaced by this value to compensate for deviations of the probe or of the measuring device Slope adjustment of sensors oFF No slope adjustment...

Page 9: ...ing tube connection for lower press in contact to ambient air by means of a tube no water contact and bring the sensor with its open pressure connection for higher pressure to water depth to be measur...

Page 10: ...l net Universal multi channel software for real time recording and displaying of measuring data with real database In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...o G1 2 pressure terminal or plug plastic tube to a suitable adapter Connection to instrument Use cable MSD K51 11 Accuracy Check Adjustment Service You can send the device to the manufacturer for adju...

Page 12: ...h to be displayed Err 4 Display range underflow Check value below 19999 Tara to low Er 11 Value could not be calculated Choose different unit Calculation overflow happened Choose different unit Err 7...

Page 13: ...ment display illuminated white Correction function offset slope via menu if offset slope value not equal zero corr is displayed Additional functions min max hold function Averaging adjustable 1 120 se...

Page 14: ...H82 0 01 6C 02 Operating Manual GMH 5130 page 14 of 14 _____________________________________________________ _____________________________________________________________________________...

Reviews: