13 Error and System Messages
Display
Description
What to do?
or
Err.9
No sensor connected
Switch off device and connect sensor
Connected sensor or device defective
If 2nd sensor available, check if device is ok. Return
defective device/sensor to manufacturer for repair
Value extremely out of measuring range
Check: pressure not within sensor range?
No display or
confused
characters,
device does not
react on keypress
Battery empty
Replace battery
Mains operation: wrong voltage or polarity
Check power supply, replace it when necessary
System error
Disconnect battery and power supplies, wait shortly,
then reconnect
Device defective
Return to manufacturer for repair
Err.1
Measured value above allowable range
Check: pressure not within sensor range?
-> measuring value to high!
Sensor defective
Return to manufacturer for repair
Err.2
Measured value below allowable range
Check: pressure not within sensor range?
-> measuring value to low!
Sensor defective
Return to manufacturer for repair
Err.3
Display range overflow
Check: value above 19999 -> to high to be displayed
Err.4
Display range underflow
Check: value below -19999 (Tara?) -> to low
Er.11
Value could not be calculated
Choose different unit
Calculation overflow happened
Choose different unit
Err.7
System error
Return to manufacturer for repair
– – – –
Sensor not present / recognised
reconnect sensor,
during logging: stop the logger and restart it
could not calculate value
suitable sensor / unit combination necessary
If
“bAt”
is flashing, the battery will be exhausted soon. Further measurements are possible for short time.
If “bAt” is displayed continuously the battery is ultimately exhausted and has to be replaced. Further
measurements aren’t possible any more.
14 Reshipment und Disposal
14.1 Reshipment
_DANGER
All devices returned to the manufacturer have to be free of any residual of measuring media
and/or other hazardous substances. Measuring residuals at housing or sensor may be a risk for
persons or environment
Use a adequate transport package for reshipment, especially for fully functional devices. Please
make sure that the device is protected in the package by enough packing materials.
14.2 Disposal instructions
Batteries must not be disposed in the regular domestic waste but at the designated collecting points.
The device must not be disposed in the unsorted municipal waste! Send the device directly to us
(sufficiently stamped), if it should be disposed. We will dispose the device appropriate and environmentally