13 Error and System Messages 

Display 

Description 

What to do? 

 

or

 

Err.9

 

No sensor connected 

Switch off device and connect sensor 

Connected sensor or device defective 

If 2nd sensor available, check if device is ok. Return 
defective device/sensor to manufacturer for repair 

Value extremely out of measuring range 

Check: pressure not within sensor range? 

No display or 

confused 

characters, 

 

device does not 

react on keypress

 

Battery empty 

Replace battery 

Mains operation: wrong voltage or polarity 

Check  power supply, replace it when necessary 

System error 

Disconnect battery and power supplies, wait shortly, 
then reconnect 

Device defective 

Return to manufacturer for repair 

Err.1 

Measured value above allowable range 

Check: pressure not within sensor range?  
-> measuring value to high! 

Sensor defective 

Return to manufacturer for repair 

Err.2 

Measured value below allowable range 

Check: pressure not within sensor range?  
-> measuring value to low! 

Sensor defective 

Return to manufacturer for repair 

Err.3 

Display range overflow 

Check: value above 19999 -> to high to be displayed 

Err.4 

Display range underflow 

Check: value below -19999 (Tara?) -> to low 

Er.11 

Value could not be calculated 

Choose different unit 

Calculation overflow happened 

Choose different unit 

Err.7 

System error 

Return to manufacturer for repair 

– – – – 

Sensor not present / recognised 

reconnect sensor,  
during logging: stop the logger and restart it 

could not calculate value 

suitable sensor / unit combination necessary 

If 

“bAt”

 is flashing, the battery will be exhausted soon. Further measurements are possible for short time.  

If “bAt” is displayed continuously the battery is ultimately exhausted and has to be replaced. Further 
measurements aren’t possible any more.  

14 Reshipment und Disposal 

14.1  Reshipment 

 

_DANGER 

        

All devices returned to the manufacturer have to be free of any residual of measuring media 
and/or other hazardous substances. Measuring residuals at housing or sensor may be a risk for 
persons or environment 

 
 
 
 

Use a adequate transport package for reshipment, especially for fully functional devices. Please 
make sure that the device is protected in the package by enough packing materials.  

 

14.2  Disposal instructions 

 

Batteries must not be disposed in the regular domestic waste but at the designated collecting points.  

The device must not be disposed in the unsorted municipal waste! Send the device directly to us 
(sufficiently stamped), if it should be disposed. We will dispose the device appropriate and environmentally 
sound.

 

 

 

Summary of Contents for GMH 5130

Page 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de WEEE Reg Nr DE 93889386 as of version 1 0 GMH 5130 Operating Manual Handheld Pressure Meter water proof Please carefully read these instructions b...

Page 2: ...LEMENTS 5 4 2 PUSHBUTTONS 5 4 3 CONNECTIONS 5 4 4 POP UP CLIP 6 5 START OPERATION 7 6 CONFIGURATION 7 7 REMARKS TO SPECIAL FEATURES 8 7 1 DIFFERENT PRESSURE UNITS 8 7 2 DIFFERENT KINDS OF MEASURING 8...

Page 3: ...to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Prot...

Page 4: ...e usage of this device at potentially explosive areas increases danger of deflagration explosion or fire due to sparking 3 Product Specification 3 1 Scope of supply The scope of supply includes GMH 51...

Page 5: ...ase refer to chapter 7 1 7 corr arrow zero point or slope correction is active 4 2 Pushbuttons On off key backlight press shortly activate backlight or switch on instrument press longer switch off ins...

Page 6: ...to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attac...

Page 7: ...es the settings enter finishes the configuration and returns to standard measuring operation Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults...

Page 8: ...sor dep e g 5 00 5 00 mbar The offset of sensor will be displaced by this value to compensate for deviations of the probe or of the measuring device Slope adjustment of sensors oFF No slope adjustment...

Page 9: ...ing tube connection for lower press in contact to ambient air by means of a tube no water contact and bring the sensor with its open pressure connection for higher pressure to water depth to be measur...

Page 10: ...l net Universal multi channel software for real time recording and displaying of measuring data with real database In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...o G1 2 pressure terminal or plug plastic tube to a suitable adapter Connection to instrument Use cable MSD K51 11 Accuracy Check Adjustment Service You can send the device to the manufacturer for adju...

Page 12: ...h to be displayed Err 4 Display range underflow Check value below 19999 Tara to low Er 11 Value could not be calculated Choose different unit Calculation overflow happened Choose different unit Err 7...

Page 13: ...ment display illuminated white Correction function offset slope via menu if offset slope value not equal zero corr is displayed Additional functions min max hold function Averaging adjustable 1 120 se...

Page 14: ...H82 0 01 6C 02 Operating Manual GMH 5130 page 14 of 14 _____________________________________________________ _____________________________________________________________________________...

Reviews: