GREISINGER GMH 3692 Operating Manual Download Page 9

H74.0.14.6C-05 

Operating Manual GMH 3692 

page 9 of 12 

7  Calibration of the Sensor 

In order to compensate for ageing of the sensor, the sensor has to be calibrated at regular intervals.  
The device is equipped with an easy-to-use calibration functions.  
We recommend to calibrate the sensor at least all 7 days, or to get maximum precision, before each 
measuring series. 

7.1  One Point Calibration ('

(AL  1-PT

') 

The calibration adjusts the sensor to the oxygen content of the atmosphere (20.95%). Therefore simply 
expose the sensor to the ambient air (sufficient ventilation in closed rooms has to be ensured) 

Start calibration: press 

-key for 2 seconds 

The display will show '

A,R

 

, and as soon the values for oxygen and temperature are stable, the 

calibration will be finished  
Then the electrode state resulting of the successful calibration will be shown for a short time (evaluation in 

10% steps:  xx% 

ELE[

). 

7.2  2 / 3-Point Calibration ('

(AL  2-PT

(AL  3-PT

') 

The sensor will be automatically calibrated to the oxygen content of the atmosphere (20.95%) and one or two 
additional concentrations. As reference gases usually Nitrogen (0% vol O

2

) or pure oxygen are used 

1.  Start calibration: press 

-key for 2 seconds 

2.  First calibration reference: (Pt.1) 

As first reference at a 3-point calibration, the zero reference has to be applied (

NULL

),  

at a 2-point calibration either 100% or 0%(

NULL

). 

The display will show 

, and the referring reference which should be applied: 

-  

NULL

 for 0% oxygen 

-  

0.2

 for pure oxygen  

As long as the display blinks, no valid reference is recognised by the instrument. 
As soon the values for oxygen and temperature are stable, the calibration of the first point will be finished

The instrument tells you to apply the next reference (possible references are blinking in the display). 

3.  Second calibration reference: (Pt.2) 

The display will show 

, and the referring reference which should be applied: 

-  

A,R

 for ambient air  

-  

0.2

 for pure oxygen  

-  

NULL

 for 0% oxygen 

As long as the display blinks, no valid reference is recognised by the instrument. 
As soon the values for oxygen and temperature are stable, the calibration of the second point will be 
finished. At 2-point calibration the calibration will be finished and the electrode state resulting of the 

successful calibration will be shown for a short time (evaluation in 10% steps:  xx% 

ELE[

). 

At 3-point calibration the instrument tells you to apply the next reference (possible reference is blinking) 

4.  Third calibration reference: (Pt.3) 

The display will show 

, and the referring reference which should be applied: 

As soon the values for oxygen and temperature are stable, the calibration of the second point will be 
finished. At 2-point calibration the calibration will be finished and the sensor state resulting of the successful 

calibration will be shown for a short time (evaluation in 10% steps:  xx% 

ELE[

). 

In case of error messages being displayed during the calibration process, please refer to our notes 
at the end of this manual! If a calibration cannot be carried out after an extended period of time, at 
least one of the measuring values is unstable (oxygen partial pressure, temperature).  
Please check your measuring arrangements! 

7.3  Evaluation of Sensor State (

ELE[

Watch sensor state:  press key "CAL" shortly once display show for a short time xx% 

ELE[

It will show the electrode state resulting of the last successful calibration carried out. 
The valuation is displayed in 10 percent steps: 100% means optimal sensor condition. Lower values are 
indicating that the sensor life time will be reached soon.  

Remark:   But also an erroneous pressure may be the cause of low valuation values. 

Summary of Contents for GMH 3692

Page 1: ...Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de As of Version 1 2 Operating Manual GMH 3692 Oxygen meter for oxygen in gases with integrated temperature and pressure m...

Page 2: ...BRATION OF THE SENSOR 9 7 1 ONE POINT CALIBRATION AL 1 PT 9 7 2 2 3 POINT CALIBRATION AL 2 PT AL 3 PT 9 7 3 EVALUATION OF SENSOR STATE ELE 9 7 4 CALIBRATION INTERVAL INT 10 8 INSPECTION OF THE ACCURAC...

Page 3: ...esigned and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reliability cannot be guaranteed unless the standard safety measures and spec...

Page 4: ...the real time clock has to be set again 2 Mains operation with power supply When using a power supply please note that operating voltage has to be 10 5 to 12 V DC Do not apply overvoltage Cheap 12V po...

Page 5: ...Signals if oxygen and temperature values are stable 5 CAL arrow Signals if automatic calibration is in progress 6 Logg arrow No function 7 Warning sign Signals weak battery or other warning message Di...

Page 6: ...points depend on current device settings To change device settings press Menu for 2 seconds This will activate the configuration menu main display Set Pressing Menu changes between the menus points p...

Page 7: ...0 C or 9 0 F 9 0 F The zero point of the temperature measuring is shifted for the entered value This can be used to compensate sensor and instrument deviations OFF No zero adjustment for temperature m...

Page 8: ...life time also suitable for application with larger CO2 concentrations GOEL 380 Fast responding for low oxygen concentration e g protection atmosphere below 1 max 25 For application without larger CO...

Page 9: ...xt reference possible references are blinking in the display 3 Second calibration reference Pt 2 The display will show and the referring reference which should be applied A R for ambient air 0 2 for p...

Page 10: ...cord and display the measuring values GMHKonfig Software for a comfortable configuration of the device e g freeware In case you want to develop your own software we offer a GMH3000 development package...

Page 11: ...ue too low Wrong sensor connected Check sensor Sensor cable or instrument defect return to manufacturer for repair ERR 7 system error return to manufacturer for repair If BAT is flashing the battery w...

Page 12: ...Break proof ABS housing Protection class Front side IP65 Dimensions L B H mm Weight Without pressure port 142 x 71 x 26 mm L x B x H approx 160 g incl battery Power supply Current consumption 9V Batt...